FDA Medical Device Recalls (openFDA)
Becton Dickinson & Co.: BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA) Catalog Number: 245128
- Recall number
- Z-2686-2024
- Recalling firm
- Becton Dickinson & Co.
- Product
- BD BACTEC MGIT 960 PZA Kit)- IVD qualitative procedure for susceptibility testing of Mycobacterium tuberculosis, from culture to pyrazinamide (PZA) Catalog Number: 245128
- Status
- Open, Classified
- Initiated
- 2024-07-18
- Posted
- 2024-08-16
- Terminated
- not on file
- Reason
- May intermittently produce falsely resistant results for pyrazinamide (PZA) during susceptibility testing of Mycobacterium tuberculosis isolates.
- Root cause
- Under Investigation by firm
- Action
- ***UPDATE 11/13/2025*** BD Issued an URGENT: Medical Device Correction letters were sent to customers. BD is pleased to announce that we have resumed production of a modified version of the BD BACTEC" MGIT" 960 PZA Kit which includes modified inoculation methods and reduced shelf-life. BD has conducted a thorough review of the performance from all supported inoculum sources using the synthetic raw material and the following change has been implemented for the BD BACTEC" MGIT" 960 PZA Kit. 1. Only inoculum prepared from MGIT tubes 3-5 days post-positivity can be currently supported. 2. Inoculum prepared from MGIT tubes 1-2 days past instrument positivity are currently not supported. 3. Inoculum prepared from solid media are currently not supported. Additionally, the shelf life of the product has been adjusted from 18 months to 13 months to reflect the latest internal supporting data. Clinical users should refer to the updated Instructions for Use (IFU) available at https://www.bd.qarad.eifu.online/hcp. Additional languages will be added as translations are completed. ***UPDATE 8/1/24*** BD issued an Urgent Medical Device Correction Update letter on 8/1/24 as lots were inadvertently missed on the communication sent on 18Jul2024. BD issued URGENT: Medical Device Correction letter on 18Jul2024 via email and mail . Letter states reason for recall, health risk and action to take: 1. Customeers should discard the affected product immediately. 2. There are no recommendations for retesting or reviewing previous patient test results. Actions to be taken by customer: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood. 3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure
- Quantity
- 12,383 eaches
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brunei, Brazil, Canada, China, Costa Rica, Dominican Republic, Guatemala, Guyana, Hong Kong, Haiti, Indonesia, India, Japan, Korea, Myanmar, Macau, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, El Salvador, Thailand, Trinidad and Tobago, Taiwan, Vietnam.
- Lot, serial or model codes
- UDI-DI: (01) 0038290245128 Lot Numbers: Amended 8/1/24 3104416 3156654 3191569 3233971 3241654 3248314 3269157 3298311 3298317 3304389 3324422 3338965 4002353 4009894 4036330 4051247
- 510(k) numbers
- ["K021582"]
- PMA numbers
- not on file
- Product code
- MJA
- Firm FEI
- 1119779
- Firm city
- Sparks
- Firm state
- MD
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28