Skip to the content

FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: iConnect Enterprise Archive. The firm name on the label is Merge Healthcare, Hartland, WI. iConnect Enterprise Archive is intended for use as a vendor neutral archive for storage and communications of medical images and data.

Recall number
Z-2686-2016
Recalling firm
Merge Healthcare, Inc.
Product
iConnect Enterprise Archive. The firm name on the label is Merge Healthcare, Hartland, WI. iConnect Enterprise Archive is intended for use as a vendor neutral archive for storage and communications of medical images and data.
Status
Terminated
Initiated
2016-01-30
Posted
not on file
Terminated
2019-10-15
Reason
Interventional Radiology (IR) images are stored as JPEG2k Lossless in Merge Enterprise Archive (EA) and are not displaying correctly in RadSuite.
Root cause
Software design
Action
The firm, Merge, sent an "URGENT: MEDICAL DEVICE RECALL" letters dated 1/29/2016 and 5/25/2016 via e-mail or via certified mail if they did not have an e-mail address for the customers. The letters described the product, problem and actions to be taken. The customers were instructed to do the following: you must discontinue using the affected version of RadSuite or iConnect Enterprise Archive with GE AW Workstation; use the workaround listed in the letter; complete and return the enclosed form no later than June 17, 2016; ensure that all users of the product are provided with the notification; if you have further distributed this product, please identify your customers and notify them at once of this product recall Merge has a released fix available for this issue. If you have any additional questions, please send an email to recall@merge.com or call 800-724-5970 ext 8027.
Quantity
96 sites potentially have the affected versions for both products
Distribution
US Distribution to: CO and OK.
Lot, serial or model codes
Version: 8.80, 8.80.0.1, 8.80.0.2, 8.80.1.1, 8.80.2, and 8.80.2.1
510(k) numbers
not on file
PMA numbers
not on file
Product code
LMB
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register