FDA Medical Device Recalls (openFDA)
Teleflex Medical: MAQUET Servo Humidifier 163, Inmed Mfg. Sdn. Bhd. Lot PT 2577, Jalan Perusahaan, 4, 34600 Kamunting, Perak, Malaysia.
- Recall number
- Z-2683-2016
- Recalling firm
- Teleflex Medical
- Product
- MAQUET Servo Humidifier 163, Inmed Mfg. Sdn. Bhd. Lot PT 2577, Jalan Perusahaan, 4, 34600 Kamunting, Perak, Malaysia.
- Status
- Terminated
- Initiated
- 2015-12-03
- Posted
- 2016-08-29
- Terminated
- 2017-01-24
- Reason
- Cracking may occur at the 15mm and 22mm ISO female connector.
- Root cause
- Device Design
- Action
- INMED sent an Urgent - Field Safety Notice dated December 3, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. They were instructed to cease use and distribution of stock of the affected product batch and quarantine immediately. Customers are to return all affected product to MAQUET. An Acknowledgement form was included to be completed and returned to their MAQUET representative. If the product was distributed, they were instructed to provide the field safety notice to all of their customers who have received product in scope of the Field Action. Their customer is then required to complete the Acknowledgement form and return it to them. Should you require any further information or support concerning this issue, please contact your local MAQUET representative at fieldactions@maquet.com. INMED is committed to providing high quality, safe and effective products. W e sincerely apologize for any inconvenience this action may cause your operations. If you have any other questions, feel free to contact your local sales representative or Customer Service. For further questions, please call (610) 378-0131.
- Quantity
- 2,700 units
- Distribution
- US Distribution to the states of : FL, KS, and WV.
- Lot, serial or model codes
- Model number: 01-06-8125-8, Product code: XKC01-06-8125-8, Lot numbers: 201450, 201451, 201452 and 201504.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BYD
- Firm FEI
- 3005747797
- Firm city
- Morrisville
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28