FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Ablation Galil Technology, IcePearl 2.1 CX 90 Cyroablation Needles, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.
- Recall number
- Z-2682-2020
- Recalling firm
- Boston Scientific Corporation
- Product
- Ablation Galil Technology, IcePearl 2.1 CX 90 Cyroablation Needles, Prescription Only Galil Medicals 2.1 CX Cryoablation Needles are intended for cryoablative destruction of tissue during surgical procedures.
- Status
- Terminated
- Initiated
- 2020-06-04
- Posted
- not on file
- Terminated
- 2024-09-23
- Reason
- The needle surface deteriorated due to unexpected electrolysis during procedure.
- Root cause
- Nonconforming Material/Component
- Action
- Boston Scientific sent a letter dated 06/04/2020 and titled, "Urgent Medical Device Removal - Immediate Action Required" to customers regarding the IceFORCE 2.1 CX and IcePearl 2.1 CX Cryoablation Needles. The firm initiated a removal of specific lots/batches of IceFORCE 2.1 CX and IcePearl 2.1 CX Cryoablation Needles. The letter states that further distribution or use of any remaining product affected by the removal should cease immediately. Customers should verify the product lot/batch number to identify whether any product within their inventory is affected. If they have affected product in their inventory customers should respond and follow the instructions on the verification form. Customers must complete the Reply Verification Tracking Form and returned even if you do not have any of the units from the recalled lot/batch and email or fax your Reply Verification Tracking Form(s) immediately to get a Returned Goods Authorization Number (RGA #) before you return any product. Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center 1-866-213-1806. Package/Ship the Recalled Product. Seal the box, and return to: Boston Scientific Corporation, US Distribution Center Boston Scientific Marina Bay Customer Fulfillment Center 500 Commander Shea Blvd. Quincy, Massachusetts 02171 RGA: Replacement product will be provided once affected product has been received by Boston Scientific Distributors should forward the notification to customers. Facilities that have sent products to another hospital within their network, should ensure that the notification is forwarded to them. Customers should contact their local sales representatives; email: BSCFieldActionCenter@bsci.com or can contact Sr. Regulatory Affairs Specialist at 508-683-40 or email: clara.johnson@bsci.com regarding any questions associated with the notification. BSCFieldActionCenter@bsci.com
- Quantity
- 205 devices
- Distribution
- Nationwide Distribution to states of: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, KS, LA, MA, MD, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI WV, and WY and the country of: Canada.
- Lot, serial or model codes
- UPN: FPRPR3617, GTIN: 07290104831072 , Batch: A0904, A1085, A1177, A1178, A6681, A6682, A6962, A6963, A6977, A7113, U0084
- 510(k) numbers
- ["K152133"]
- PMA numbers
- not on file
- Product code
- GEH
- Firm FEI
- 3005099803
- Firm city
- Marlborough
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28