Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Electronics North America Corporation: Philips DuraDiagnost stationary X-ray system

Recall number
Z-2682-2016
Recalling firm
Philips Electronics North America Corporation
Product
Philips DuraDiagnost stationary X-ray system
Status
Terminated
Initiated
2015-09-17
Posted
2016-09-01
Terminated
2017-08-31
Reason
Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].
Root cause
Radiation Control for Health and Safety Act
Action
Philips plans to bring the products into compliance with the Federal standard at no cost to customers by implementing a field correction. 1. You will contact customers and initiate the actions to correct the failure to comply issues. 2. You will attach the warning label to maintain an audible volume, and issue a supplement to the user manual that instructs the user how to adjust the volume for the local noise environment and include a warning that the user shall maintain an audible volume. 3. You will attach the identification label to the affected Height-Adjustable Tabletop. 4. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. If you need any further information or support concerning this, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
Quantity
15
Distribution
US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.
Lot, serial or model codes
SN: 140054 140059 150013 150014 150039 150041 150042 150052 150053 150055 150056 150059 150061 150091 150093
510(k) numbers
["K141381"]
PMA numbers
not on file
Product code
KPR
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register