Philips Electronics North America Corporation: Philips DuraDiagnost stationary X-ray system
- Recall number
- Z-2682-2016
- Recalling firm
- Philips Electronics North America Corporation
- Product
- Philips DuraDiagnost stationary X-ray system
- Status
- Terminated
- Initiated
- 2015-09-17
- Posted
- 2016-09-01
- Terminated
- 2017-08-31
- Reason
- Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels [21 CFR 1010.2., 21 CFR 1010.3(a), and 21 CFR 1020.30(3)].
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Philips plans to bring the products into compliance with the Federal standard at no cost to customers by implementing a field correction. 1. You will contact customers and initiate the actions to correct the failure to comply issues. 2. You will attach the warning label to maintain an audible volume, and issue a supplement to the user manual that instructs the user how to adjust the volume for the local noise environment and include a warning that the user shall maintain an audible volume. 3. You will attach the identification label to the affected Height-Adjustable Tabletop. 4. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. If you need any further information or support concerning this, please contact your local Philips representative or Philips Customer Service: 1-800-722-9377.
- Quantity
- 15
- Distribution
- US Distribution to the states of : ME, NE, OR, CA, TX, NC, WA and OH.
- Lot, serial or model codes
- SN: 140054 140059 150013 150014 150039 150041 150042 150052 150053 150055 150056 150059 150061 150091 150093
- 510(k) numbers
- ["K141381"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28