Skip to the content

FDA Medical Device Recalls (openFDA)

Nanosphere, Inc.: The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

Recall number
Z-2682-2014
Recalling firm
Nanosphere, Inc.
Product
The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.
Status
Terminated
Initiated
2013-10-10
Posted
2014-09-23
Terminated
2015-06-16
Reason
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Root cause
Nonconforming Material/Component
Action
Nanosphere sent a Field Notification Letter dated October 10, 2013, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Instructions were included in the letter for customers to determine whether or not the extraction tray lots should be returned. If the extraction tray lots are to be returned, customers were to complete and return the attached form and replacement tip holder assemblies will be supplied at no cost. Customers could then dispose of the recalled unused tip holder assemblies. Customers with questions were instructed to contact their Nanosphere technical support representative. For questions regarding this recall call 1-888-837-4436, ext 3.
Quantity
80 units
Distribution
Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
Lot, serial or model codes
Product Number: 20-009-022; Lot Number and Expiration Date: Lot: 100313022B, Expiration Date: 04/01/14
510(k) numbers
["K123197"]
PMA numbers
not on file
Product code
OZN
Firm FEI
3006028115
Firm city
Northbrook
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register