FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
- Recall number
- Z-2680-2014
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- The Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of the whole Body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients.
- Status
- Terminated
- Initiated
- 2014-08-27
- Posted
- 2014-09-23
- Terminated
- 2015-03-30
- Reason
- Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest, the examination may need to be repeated.
- Root cause
- Under Investigation by firm
- Action
- The firm, Siemens, sent a "Safety Advisory Notice" dated August 27, 2014, to end users/customers. The notice identified the product, problem, and action to be taken. The customers were instructed to follow the instructions in the notice; promptly notify and instruct all the staff at your organization who have to be aware of this problem; forward this safety information to other organizations affected by this measure; and if the device has been sold and therefore no longer in your possession, forward to new owner. Siemens prepared a field modification of the Ysio systems with a hardware upgrade. This field modification will be available in September 2014. If you have any questions, please contact Regulatory Specialist at 610-219-4334 or Customer support at 610-219-6300.
- Quantity
- 62
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Model numbers: 10281013, 10281163.
- 510(k) numbers
- ["K081722"]
- PMA numbers
- not on file
- Product code
- KPR
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28