Skip to the content

FDA Medical Device Recalls (openFDA)

Remel Inc: Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is a selective and differential chromogenic medium recommended for use in the qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA in healthcare settings. It is also intended for use in the qualitative detection of MRSA from positive blood cultures demonstrating gram-positive cocci on Gram stain.

Recall number
Z-2677-2011
Recalling firm
Remel Inc
Product
Remel Spectra MRSA, Ref R01821, 10 plates per package, and Ref R01822, 100 plates per package. The firm name on the label is Remel, Lenexa, KS. The product is a selective and differential chromogenic medium recommended for use in the qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA in healthcare settings. It is also intended for use in the qualitative detection of MRSA from positive blood cultures demonstrating gram-positive cocci on Gram stain.
Status
Terminated
Initiated
2011-05-11
Posted
2011-06-28
Terminated
2013-06-18
Reason
The product failed to prevent growth of methicillin-sensitive Staphylococcus aureus (MSSA)
Root cause
Process control
Action
Remel, Inc. sent an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" letter, dated May 10, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to review their inventory and discard all remaining product lots. Additionally, customers are to complete the Product Inventory Checklist and indicate whether or not they desire replacement of the product or credit. A self-addressed , stamped envelope was enclosed in the letter to use to return the checklist. Questions can be answered by contacting the Technical Services Department at 800-255-6730 or 913-888-0939.
Quantity
187/10-packs and 261/100-packs
Distribution
Nationwide Distribution--including Puerto Rico.
Lot, serial or model codes
Cat. # R01821 (10 pack) - Lot 028251, exp. 2011.06.13;  Cat. #R01822 (100 pack) - Lot 028252, exp. 2011.06.13; 028883, exp. 2011.06.14; 028884, exp. 2011.06.14; and 031922, exp. 2011.06.21
510(k) numbers
["K092407"]
PMA numbers
not on file
Product code
JSO
Firm FEI
1924669
Firm city
Lenexa
Firm state
KS
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register