FDA Medical Device Recalls (openFDA)

Intuitive Surgical, Inc.: STAPLER 30,IS4000; Model number 470430; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.

Recall number
Z-2676-2016
Recalling firm
Intuitive Surgical, Inc.
Product
STAPLER 30,IS4000; Model number 470430; General and Plastic Surgery: The Intuitive Surgical da Vinci EndoWrist Stapler 30 Instrument and Stapler 30 Reloads are intended to be used with the da Vinci Xi Surgical System (Model IS4000) for resection, transection and/or creation of anastomoses in General, Thoracic, Gynecologic, and Urologic surgery. The device is indicated for adult and pediatric use. The device can be used with staple line or tissue buttressing material.
Status
Terminated
Initiated
2016-07-14
Posted
2016-08-26
Terminated
2017-01-03
Reason
Field failures were reported related to the da Vinci Xi Surgical System EndoWrist Stapler 45 and 30 instruments having an interruption in shaft rotation and / or instrument jaws not opening.
Root cause
Nonconforming Material/Component
Action
Intuitive sent a Stop Use e-mail to all affected sites on July 14, 2016 to inform users of the issue prior to the commencement of a recall. E-mail described the problem and asked that all surgeons who use the stapler be informed to immediately stop use. Formal letters will follow. Medical Device Recall letters were sent on July 21, 2016. Customers were instructed to return quarantined affected product whether used or unused to Intuitive Surgical. Customers should contact customer service to initiate the standard Return Material Authorization (RMA) process by phone or email: North and South America: 800-876-1310, option 3 (6AM to 5PM PST) Email: customersupport..servicesupport@intusurg.com Europe, Middle East, Asia and Africa Phone: 800-821-2020 or +41 21 821 2020 Email: ics@intusurg.com Customers were also instructed to complete the attached Return Acknowledgement Form and submit via email to Regulatory Compliance at isi.compliance@intusurg.com or by fax to +1 (408) 523-0619. For questions regarding this recall call 408-523-2100. 10-31-16 Update: Monthly status update states that an additional 93 instruments have been identified and an additional communication will be sent to these sites in Mid November 2016,
Quantity
875 instruments (827 a la carte + 48 in kits) total, all models
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Belgium, Denmark, France, Germany. Italy. Japan, Spain, Sweden, Switzerland, Taiwan, and Turkey,
Lot, serial or model codes
All model 470430
510(k) numbers
["K140553"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register