Peters Surgical: The synthetic non-absorbable surgical sutures CARDIOFLON Evolution are composed of polyester [poly(ethylene terephtalate)] braids, coated, with or without PLEDGETS of different sizes. CARDIOFLON Evolution sutures are green dyed (D&C green # 6, CI 61565) and white (undyed). Some packs are composed of green and white braids to facilitate location during use. CARDIOFLON Evolution sutures comply with the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP) for non-absorbable surgical sutures, except for some diameters. Type of packaging: - Unit: Sutures are packed in a support card Onepak or in a pack under an external polyethylene/ paper peel pouch. - Box: 36 units for single or double armed (36 sutures), 24 units of ligatures (24 reels), 12 packs for double armed (10 Green or 5 Green/5 White) (120 sutures).

Recall number
Z-2676-2014
Recalling firm
Peters Surgical
Product
The synthetic non-absorbable surgical sutures CARDIOFLON Evolution are composed of polyester [poly(ethylene terephtalate)] braids, coated, with or without PLEDGETS of different sizes. CARDIOFLON Evolution sutures are green dyed (D&C green # 6, CI 61565) and white (undyed). Some packs are composed of green and white braids to facilitate location during use. CARDIOFLON Evolution sutures comply with the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP) for non-absorbable surgical sutures, except for some diameters. Type of packaging: - Unit: Sutures are packed in a support card Onepak or in a pack under an external polyethylene/ paper peel pouch. - Box: 36 units for single or double armed (36 sutures), 24 units of ligatures (24 reels), 12 packs for double armed (10 Green or 5 Green/5 White) (120 sutures).
Status
Terminated
Initiated
2014-06-25
Posted
2014-09-22
Terminated
2017-12-13
Reason
A potential splitting of the surface treatment of the thread ends on certain sutures has recently been identified.
Root cause
Process control
Action
The sole US distributor (MedAlliance Inc.) was notified of the recall via e-mail on July 16, 2014. IMPORTANT AND URGENT - BATCH RECALL PROCEDURE letters dated June 25, 2014 under the Peters Surgical header were sent to all MedAlliance Inc. consignees. The letters included instructions for consignees to contact MedAlliance Inc. by phone (630-933-8661) or by e-mail (agoos@medalliancesolutions.com) for information on what to do with the recalled products. Consignees were also notified that they can contact Peters Surgical directly at qualite@peters-surgical.com for answers to any additional questions. MedAlliance Inc. sent VOLUNTARY PRODUCT RECALL/WITHDRAWAL LETTER[s] dated June 25, 2014 to all consignees. The letters included instructions for consignees to: 1) discontinue use of and segregate recalled product in a secure location for return to MedAlliance Inc., and 2) to complete and return the enclosed RETURNED GOOD AUTHORIZATION (RGA) FORM to MedAlliance Inc. to arrange for the return of the recalled products. Consignees with any questions about the recall can contact Ms. Alaina Goos, Senior Product Manager, by phone at 630-933-8661 x 104 or via e-mail at agoos@medalliancesolutions.com.
Quantity
22,190 units
Distribution
Worldwide Distribution- US: Nationwide (CA, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NY, OH, OR, PA, SC, TX, VA, WA); and countries: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CAMBODIA, CHINA, CYPRUS, CZECH REPUBLIC, DOMINICAN REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, LEBANON, LIBYA, MALAYSIA, MAURITIUS, MOLDOVA, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NORTH KOREA, PAKISTAN, PERU, POLAND, ROMANIA, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, VENEZUELA, VIETNAM, YEMEN
Lot, serial or model codes
1) Item reference: 19P20Q , Batch number: 1135 , Expiration date: 10/2018 , GTIN on box label: 3661522034989 ; 2) Item reference: 19P30AC , Batch number: 942720 , Expiration date: 03/2018 , GTIN on box label: 3661522034446 ; 3) Item reference: 19P30AE , Batch number: 7627 , Expiration date: 04/2019 , GTIN on box label: 3661522034477 ; 4) Item reference: 19P30AE , Batch number: 942910 , Expiration date: 02/2018 , GTIN on box label: 3661522034477 ; 5) Item reference: 19P30AE , Batch number: 971670 , Expiration date: 05/2018 , GTIN on box label: 3661522034477 ; 6) Item reference: 19P30AK , Batch number: 942530 , Expiration date: 03/2018 , GTIN on box label: 3661522034552 ; 7) Item reference: 19P30AK , Batch number: 978580 , Expiration date: 06/2018 , GTIN on box label: 3661522034552 ; 8) Item reference: 19P30AN , Batch number: 942500 , Expiration date: 02/2018 , GTIN on box label: 3661522034576 ; 9) Item reference: 19P30AQ , Batch number: 2480 , Expiration date: 10/2018 , GTIN on box label: 3661522034606 ; 10) Item reference: 19P30AQ , Batch number: 5508 , Expiration date: 02/2019 , GTIN on box label: 3661522034606 ; 11) Item reference: 19P30AQ , Batch number: 994990 , Expiration date: 07/2018 , GTIN on box label: 3661522034606 ; 12) Item reference: 19P30AQ , Batch number: 1008710 , Expiration date: 08/2018 , GTIN on box label: 3661522034606 ; 13) Item reference: 19P30N , Batch number: 942420 , Expiration date: 02/2018 , GTIN on box label: 3661522034262 ; 14) Item reference: 19P30R , Batch number: 4370 , Expiration date: 01/2019 , GTIN on box label: 3661522034279 ; 15) Item reference: 19P30R , Batch number: 5021 , Expiration date: 01/2019 , GTIN on box label: 3661522034279 ; 16) Item reference: 19P30R , Batch number: 5557 , Expiration date: 02/2019 , GTIN on box label: 3661522034279 ; 17) Item reference: 19P30R , Batch number: 948700 , Expiration date: 03/2018 , GTIN on box label: 3661522034279 ; 18) Item reference: 19P30R , Batch number: 978410 , Expiration date: 06/2018 , GTIN on box label: 3661522034279 ; 19) Item reference: 19P30S , Batch number: 994580 , Expiration date: 07/2018 , GTIN on box label: 3661522034323 ; 20) Item reference: 19P30T , Batch number: 4371 , Expiration date: 01/2019 , GTIN on box label: 3661522034347 ; 21) Item reference: 19P30T , Batch number: 4974 , Expiration date: 02/2019 , GTIN on box label: 3661522034347 ; 22) Item reference: 19P30T , Batch number: 948740 , Expiration date: 03/2018 , GTIN on box label: 3661522034347 ; 23) Item reference: 19P30U , Batch number: 948770 , Expiration date: 03/2018 , GTIN on box label: 3661522034354 ; 24) Item reference: 19P30V , Batch number: 4372 , Expiration date: 01/2019 , GTIN on box label: 3661522034286 ; 25) Item reference: 19P30V , Batch number: 948790 , Expiration date: 03/2018 , GTIN on box label: 3661522034286 ; 26) Item reference: 19P30V , Batch number: 992210 , Expiration date: 07/2018 , GTIN on box label: 3661522034286 ; 27) Item reference: 19P30X , Batch number: 967150 , Expiration date: 05/2018 , GTIN on box label: 3661522034392 ; 28) Item reference: 19P30X , Batch number: 999390 , Expiration date: 09/2018 , GTIN on box label: 3661522034392 ; 29) Item reference: 19S15A , Batch number: 2192 , Expiration date: 10/2018 , GTIN on box label: 3661522035009 ; 30) Item reference: 19S15G , Batch number: 2195 , Expiration date: 10/2018 , GTIN on box label: 3661522035061 ; 31) Item reference: 19S15G , Batch number: 2440 , Expiration date: 10/2018 , GTIN on box label: 3661522035061 ; 32) Item reference: 19S15G , Batch number: 942960 , Expiration date: 02/2018 , GTIN on box label: 3661522035061 ; 33) Item reference: 19S15G , Batch number: 970060 , Expiration date: 05/2018 , GTIN on box label: 3661522035061 ; 34) Item reference: 19S15G , Batch number: 982470 , Expiration date: 06/2018 , GTIN on box label: 3661522035061 ; 35) Item reference: 19S15G , Batch number: 994930 , Expiration date: 07/2018 , GTIN on box label: 3661522035061 ; 36) Item reference: 19S20J , Batch number: 942940 , Expiration date: 05/2018 , GTIN on box label: 3661522034910 ; 37) Item reference: 19S20J , Batch number: 966100 , Expiration date: 08/2018 , GTIN on box label: 3661522034910 ; 38) Item reference: 19S20R , Batch number: 1872 , Expiration date: 09/2018 , GTIN on box label: 3661522037539 ; 39) Item reference: 19S30AB , Batch number: 7057 , Expiration date: 04/2019 , GTIN on box label: 3661522034439 ; 40) Item reference: 19S30AB , Batch number: 991250 , Expiration date: 07/2018 , GTIN on box label: 3661522034439 ; 41) Item reference: 19S30AB , Batch number: 999910 , Expiration date: 08/2018 , GTIN on box label: 3661522034439 ; 42) Item reference: 19S30AE , Batch number: 942400 , Expiration date: 03/2018 , GTIN on box label: 3661522034460 ; 43) Item reference: 19S30AE , Batch number: 984600 , Expiration date: 06/2018 , GTIN on box label: 3661522034460 ; 44) Item reference: 19S30AK , Batch number: 942510 , Expiration date: 02/2018 , GTIN on box label: 3661522034538 ; 45) Item reference: 19S30AK , Batch number: 992600 , Expiration date: 07/2018 , GTIN on box label: 3661522034538 ; 46) Item reference: 19S30AK , Batch number: 1011370 , Expiration date: 09/2018 , GTIN on box label: 3661522034538 ; 47) Item reference: 19S30AQ , Batch number: 942580 , Expiration date: 03/2018 , GTIN on box label: 3661522034590 ; 48) Item reference: 19S30AW , Batch number: 1485 , Expiration date: 09/2018 , GTIN on box label: 3661522037492 ; 49) Item reference: 19S30AX , Batch number: 1486 , Expiration date: 09/2018 , GTIN on box label: 3661522037508 ; 50) Item reference: 19S30AY , Batch number: 1487 , Expiration date: 09/2018 , GTIN on box label: 3661522037515 ; 51) Item reference: 19S30B , Batch number: 3403 , Expiration date: 01/2019 , GTIN on box label: 3661522034118 ; 52) Item reference: 19S30B , Batch number: 5136 , Expiration date: 02/2019 , GTIN on box label: 3661522034118 ; 53) Item reference: 19S30B , Batch number: 6222 , Expiration date: 02/2019 , GTIN on box label: 3661522034118 ; 54) Item reference: 19S30B , Batch number: 942350 , Expiration date: 03/2018 , GTIN on box label: 3661522034118 ; 55) Item reference: 19S30B , Batch number: 1009510 , Expiration date: 08/2018 , GTIN on box label: 3661522034118 ; 56) Item reference: 19S30C , Batch number: 3995 , Expiration date: 01/2019 , GTIN on box label: 3661522034125 ; 57) Item reference: 19S30C , Batch number: 4920 , Expiration date: 01/2019 , GTIN on box label: 3661522034125 ; 58) Item reference: 19S30C , Batch number: 5158 , Expiration date: 02/2019 , GTIN on box label: 3661522034125 ; 59) Item reference: 19S30C , Batch number: 6554 , Expiration date: 03/2019 , GTIN on box label: 3661522034125 ; 60) Item reference: 19S30C , Batch number: 977760 , Expiration date: 06/2018 , GTIN on box label: 3661522034125 ; 61) Item reference: 19S30C , Batch number: 988910 , Expiration date: 07/2018 , GTIN on box label: 3661522034125 ; 62) Item reference: 19S30Z , Batch number: 942670 , Expiration date: 03/2018 , GTIN on box label: 3661522034415 ; 63) Item reference: 19S35A , Batch number: 942760 , Expiration date: 04/2018 , GTIN on box label: 3661522034675
510(k) numbers
["K060163"]
PMA numbers
not on file
Product code
GAT
Firm FEI
3004060107
Firm city
Bobigny
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register