Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Surgical Inc: Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00 Software Version: N/A Product Description: Single Sterile Extension Hose with PLC Connector Component: N/A

Recall number
Z-2674-2026
Recalling firm
Zimmer Surgical Inc
Product
Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Single Hose with CPC Connectors Model/Catalog Number: 60-4009-101-00 Software Version: N/A Product Description: Single Sterile Extension Hose with PLC Connector Component: N/A
Status
Open, Classified
Initiated
2026-06-09
Posted
2026-07-02
Terminated
not on file
Reason
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.
Root cause
Under Investigation by firm
Action
On June 8, 2026, the firm began notifying consignees titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to locate and quarantine affected product in their inventory. Customers may either return the affected product, or a Zimmer Biomet sales representative may remove and return the affected product on the customer's behalf. If the product has been further distributed, the distributor should notify downstream customers and accounts of the recall. All customers should ensure that affected personnel are aware of the recall. This recall is related to prior recall RES 98665, conducted to remove Mixing Bowl and Spatula due to similar issues with seal integrity.
Quantity
14360
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and EMEA.
Lot, serial or model codes
Lot Code: Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271022(10)79701946 ; Box (10-Pack) UDI (01)00889024380455(17)271022(10)79701946 ; Individual UDI (01)00889024376786(17)271022(10)79701946 ; Lot Number 79701946 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271028(10)79701947 ; Box (10-Pack) UDI (01)00889024380455(17)271028(10)79701947 ; Individual UDI (01)00889024376786(17)271028(10)79701947 ; Lot Number 79701947 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)271030(10)79701948 ; Box (10-Pack) UDI (01)00889024380455(17)271030(10)79701948 ; Individual UDI (01)00889024376786(17)271030(10)79701948 ; Lot Number 79701948 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280325(10)80555938 ; Box (10-Pack) UDI (01)00889024380455(17)280325(10)80555938 ; Individual UDI (01)00889024376786(17)280325(10)80555938 ; Lot Number 80555938 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280402(10)80555939 ; Box (10-Pack) UDI (01)00889024380455(17)280402(10)80555939 ; Individual UDI (01)00889024376786(17)280402(10)80555939 ; Lot Number 80555939 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280326(10)80555940 ; Box (10-Pack) UDI (01)00889024380455(17)280326(10)80555940 ; Individual UDI (01)00889024376786(17)280326(10)80555940 ; Lot Number 80555940 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280911(10)80581869 ; Box (10-Pack) UDI (01)00889024380455(17)280911(10)80581869 ; Individual UDI (01)00889024376786(17)280911(10)80581869 ; Lot Number 80581869 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280331(10)80581870 ; Box (10-Pack) UDI (01)00889024380455(17)280331(10)80581870 ; Individual UDI (01)00889024376786(17)280331(10)80581870 ; Lot Number 80581870 Model No. 60-4009-101-00 ; Case (20-pack) UDI (01)00889024382336(17)280402(10)80765021 ; Box (10-Pack) UDI (01)00889024380455(17)280402(10)80765021 ; Individual UDI (01)00889024376786(17)280402(10)80765021 ; Lot Number 80765021
510(k) numbers
not on file
PMA numbers
not on file
Product code
KCY
Firm FEI
1526350
Firm city
Dover
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register