Skip to the content

FDA Medical Device Recalls (openFDA)

MedicalCommunications GmbH: Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)

Recall number
Z-2673-2025
Recalling firm
MedicalCommunications GmbH
Product
Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System)
Status
Open, Classified
Initiated
2025-09-09
Posted
2025-09-30
Terminated
not on file
Reason
Potential that the measured value may be smaller than the actual area.
Root cause
Software design
Action
Ashvins notified consignees on about 09/09/2025 via emailed letter. Consignees were instructed to cease using the freeform measurement function in the MMV until the software update has been performed, which is now available. Consignees were instructed to perform the update as soon as possible, complete the digital Field Safety Notice Acknowledgement and Receipt Form, and maintain a copy of the letter and the user manual with the device until the software update is complete.
Quantity
474 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model Name: Ashvins variant HEYEX 2 / HEYEX PACS. UDI-DI: 4260648620046. Software Version: 2.6.0 (Build 2088) to 2.6.8 (Build 2220).
510(k) numbers
K200132
PMA numbers
not on file
Product code
NFJ
Firm FEI
3003995303
Firm city
Heidelberg
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register