Instrumentation Laboratory Co.: Instrumentation Laboratory GEM Premier 3500 Blood Gas Analyzer PN 00026000000 The GEM Premier 3500 blood gas analyzer is a portable system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting. The instrument provides quantitative measurements of whole blood for pH, pC02, p02, Na+, K+, Ca++, glucose, lactate and hematocrit. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, oxygen delivery capacity, and electrolyte and metabolite balance.

Recall number
Z-2673-2011
Recalling firm
Instrumentation Laboratory Co.
Product
Instrumentation Laboratory GEM Premier 3500 Blood Gas Analyzer PN 00026000000 The GEM Premier 3500 blood gas analyzer is a portable system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting. The instrument provides quantitative measurements of whole blood for pH, pC02, p02, Na+, K+, Ca++, glucose, lactate and hematocrit. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, oxygen delivery capacity, and electrolyte and metabolite balance.
Status
Terminated
Initiated
2011-05-11
Posted
2011-06-27
Terminated
2013-03-15
Reason
Software upgrade: misuse of the barcode gun by repeatedly clicking can cause the instrument to attach test results to an incorrect patient ID.
Root cause
Software design
Action
Instrumentation Laboratories sent a "FIELD SAFETY NOTIFICATION" letter dated April 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. Included with the letter was an upgrade package containing Software V7.2.2 disk along with installation instructions. Also included with the upgrade kit was a Upgrade Tracking Form to be completed and returned via fax at 781-861-4207 or via e-mail at ra-usa@ilww.com. Customers with questions were directed to contact their local representative.
Quantity
840 units
Distribution
Worldwide Distribution-- USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE , INDONESIA, ISRAEL, ITALY, JAPAN , KOREA, REPUBLIC OF CHINA, MALAYSIA, NETHERLANDS , PHILIPPINES, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, THAILAND ,TURKEY, and UNITED KINGDOM.
Lot, serial or model codes
Serial Numbers range: 09020100 through 11041157.
510(k) numbers
["K052121"]
PMA numbers
not on file
Product code
CHL
Firm FEI
1217183
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register