FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD20/10 biplane, 722029; UNIQ FD10 OR Table, 722033; UNIQ FD10/10 OR Table, 722034; UNIQ FD20 OR Table, 722035; UNIQ FD20/20, 722038; UNIQ FD20/20 biplane OR Table, 722039; UNIQ FD20/15, 722058; UNIQ FD20/15 OR Table 722059
- Recall number
- Z-2670-2016
- Recalling firm
- Philips Electronics North America Corporation
- Product
- UNIQ 1.0: UNIQ FD10, 722026; UNIQ FD10/10, 722027; UNIQ FD20, 722028; UNIQ FD20/10 biplane, 722029; UNIQ FD10 OR Table, 722033; UNIQ FD10/10 OR Table, 722034; UNIQ FD20 OR Table, 722035; UNIQ FD20/20, 722038; UNIQ FD20/20 biplane OR Table, 722039; UNIQ FD20/15, 722058; UNIQ FD20/15 OR Table 722059
- Status
- Terminated
- Initiated
- 2016-07-07
- Posted
- not on file
- Terminated
- 2020-07-14
- Reason
- Increase in the failure rate of certain low-voltage DC power supplies (DCPS)
- Root cause
- Component change control
- Action
- Philips sent an Urgent - Field Safety Notice dated June 20, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Field Service Engineers will replace the affected DCPS free of charge. If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. For further questions regarding this recall please call (978) 659-3000.
- Quantity
- 1536 devices (includes both Allura and UNIQ)
- Distribution
- Nationwide, Canada, Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Bulgaria CHILE China Colombia Croatia Cuba Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Finland France French Polynesia Georgia Germany Greece Honduras Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Lebanon Lithuania Macedonia Malaysia Malta Mexico Morocco Myanmar Nepal Netherlands New Caledonia Norway Oman Pakistan Palestine, State of PERU Philippines Poland Portugal Qatar R¿union Romania RUSSIAN FEDERAT Russian Federation Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tanzania Thailand Tunisia Turkey Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Viet Nam
- Lot, serial or model codes
- UNIQ with release R1.0
- 510(k) numbers
- ["K141979"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28