Skip to the content

FDA Medical Device Recalls (openFDA)

Laborie Medical Technologies Corp: Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);

Recall number
Z-2669-2026
Recalling firm
Laborie Medical Technologies Corp
Product
Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);
Status
Open, Classified
Initiated
2026-05-21
Posted
2026-07-02
Terminated
not on file
Reason
Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.
Root cause
Device Design
Action
Laborie Medical Technologies notified consignees on about 05/21/2026 via letter. Consignees were instructed to always use the pre-check instructions provided in the IFU to ensure that the device is operating properly, ensure all operators are aware of the outlined IFU updates in the customer notification to include the following: 1. Updated 3-Minute Blow-Out Moisture Removal Maintenance Procedure (New IFU Section 3.0 and provided in the attached insert). 2. Laborie has added the following to the revised IFU in addition to the blow-out procedure presented in Attachment 1: " A Device Use Life Statement in Section 1.8. " The statement in Section 2.4.3. Consignees were also instructed to review the Enclosed IFU Insert Section 3.0 for the updated Blow-Out Moisture Removal Maintenance Procedure and ensure all clinical users are trained to the new maintenance procedure. Laborie recommends that customers print-out the Blow-Out procedure and post it on or near the device for awareness of the change in the Blow-Out Cycle. Distribute this notification and IFU insert to all clinical operators, supervisors and trainers, biomedical engineering / technical support staff, and any personnel responsible for system cleaning, setup, or maintenance.
Quantity
372 units
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Namibia, Nepal, Netherlands. Oman, Osterreich, Pakistan, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, United Kingdom, and Vietnam.
Lot, serial or model codes
1. Catalog Number: G3-6; UDI-DI: Primary UDI-DI: 08717775955047; Basic UDI-DI: 87177759500G3G8; All serial numbers; 2. Catalog Number: G3-7; UDI-DI: Primary UDI-DI: 08717775955054; Basic UDI-DI: 87177759500G3G8; All serial numbers; 3. Catalog Number: G3-8; UDI-DI: Primary UDI-DI: 08717775955061; Basic UDI-DI: 87177759500G3G8; All serial numbers; 4. Catalog Number: G3-12; UDI-DI: Primary UDI-DI: 08717775955078; Basic UDI-DI: 87177759500G3G8; All serial numbers; 5. Catalog Number: G3-14; UDI-DI: Primary UDI-DI: 08717775955085; Basic UDI-DI: 87177759500G3G8; All serial numbers;
510(k) numbers
["K121014"]
PMA numbers
not on file
Product code
FFX
Firm FEI
3017448360
Firm city
Portsmouth
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register