Olympus Corporation of the Americas: Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354 INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba oint" INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point INS-5017 17ga x 155mm Tip Tracked Bx Needle INS-5016 17ga x 155mm Tip Tracked Bx Needle

Recall number
Z-2663-2023
Recalling firm
Olympus Corporation of the Americas
Product
Veran : Percutaneous Always on Track Models: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354 INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba oint" INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point INS-5017 17ga x 155mm Tip Tracked Bx Needle INS-5016 17ga x 155mm Tip Tracked Bx Needle
Status
Open, Classified
Initiated
2023-08-10
Posted
2023-09-29
Terminated
not on file
Reason
Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment
Root cause
Under Investigation by firm
Action
Olympus Corporation of the Americas ( Olympus ),[Veran Medical Technologies, a wholly owned subsidiary] issued Urgent Medical Device Recall Letter on 8/10/23 to: Endoscopy Department, Pulmonary Department, Risk Management for the VERAN ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments. Letter states reason for recall, health risk and action to take: Veran requests you to take the following actions: 1. Inspect your inventory and identify any Veran models. Please check all areas of the hospital to determine if any of these devices remain in inventory. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge. Your Olympus Customer Service representative will advise you on reimbursement for your Veran devices. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0430 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at Tara.Safi@Olympus.com.
Quantity
522 units
Distribution
Worldwide distribution - US Nationwide and the countries of CN, DE, HK, IT, KR, SG, TH, TW.
Lot, serial or model codes
Model-UDI: INS-5040 Always-On vTrack Universal Tracker, LG 7-12ga 00815686020705; INS-5039 Always-On vTrack Universal Tracker, SM/CL, 12-18ga 00815686020699; INS-5036 "22 UTW 125mm Always-On Tip Tracked Biopsy Needle, Trocar point" 00815686020361; INS-5034 17 UTW Lg 123mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020354; INS-5032 "19 UTW 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" 00815686020330; INS-5029 "19 UTW 105 mm Always-On Tip Tracked biopsy needle, Chiba point" 00815686020309; INS-5028 19 UTW 155 mm Always-On Tip Tracked biopsy needle, Chiba point 00815686020293; INS-5024 "17ga 150mm Always-On Tip Tracked Biopsy Needle, Chiba point" 00815686020255; INS-5023 17ga 100mm Always-On Tip Tracked Biopsy Needle, Chiba point 00815686020248; INS-5017 17ga x 155mm Tip Tracked Bx Needle 00815686020224; INS-5016 17ga x 155mm Tip Tracked Bx Needle 00815686020217; All serial numbers and all lot numbers
510(k) numbers
["K093146","K122106","K170023"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2429304
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register