Impac Medical Systems Inc: Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.

Recall number
Z-2662-2011
Recalling firm
Impac Medical Systems Inc
Product
Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings.
Status
Terminated
Initiated
2011-03-10
Posted
2011-06-24
Terminated
2011-09-16
Reason
There's a problem in the database conversion program that is used when upgrading from the Multi-ACCS to MOSAIQ or from one version of MOSAIQ to another. If the gantry start angle is incorrect prior to pressing beam-on, the patient will be mistreated.
Root cause
Software design (manufacturing process)
Action
The firm, Elekta, sent a "USER NOTICE" to its customers with an attached confirmation of receipt form. The notice describes the product, problem and actions to be taken. The customers were instructed to follow the Interim mitigation instructions noted in the notice -verify the Arc start angle of the Simple ARC with MLC treatment field definition prior to delivery and if the customer identifes a field with an errant gantry start angle, to cease using that plan immediately and consult with the original plan for that field and make the correction to the gantry start angle. In addition, the customers were instructed to complete and return the Confirmation of Receipt via fax to: North America and Asia at 702-992-5002 or Europe and Africa +44 (0)1293 654401, Attn: Elekta OIS Su[pport and/or scan and email to: support@impact.com, europe.support@impact.com or by mail; and distribute this notice to any all users of IMPAC software at their organization who are potentially affected by this issue. For questions or requests for upgrade, email support@impac.com (North America/Asia) or europe.support@impac.com (Europe) or call USA 1-800-488-4672.
Quantity
840
Distribution
Worldwide distribution: USA (nationwide) and country of Canada.
Lot, serial or model codes
MOSAIQ 1.30, 1.60, 2.00, 2.10, and 2.20; including all service packs.
510(k) numbers
["K981313"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1000123805
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register