FDA Medical Device Recalls (openFDA)

Leica Microsystems, Inc.: Nussloch GmbH ASP6025 Tissue Processor

Recall number
Z-2660-2017
Recalling firm
Leica Microsystems, Inc.
Product
Nussloch GmbH ASP6025 Tissue Processor
Status
Open, Classified
Initiated
2017-06-19
Posted
not on file
Terminated
not on file
Reason
Incorrect labeling for specified voltage for the Alarm Connectors on the rear side of the unit.
Root cause
Labeling mix-ups
Action
Customers were notified via email on 06/19/2017. Instructions to customers included to clean surface around the Alarm Connectors and place the newly provided label at the Alarm Connectors (over the engraved specification). They were also instructed to update their current Instructions For Use and update the online help function in the software, as well as return the response form.
Quantity
812 units
Distribution
Domestic: AR, AZ, CA, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV, & DC. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, England, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lithuania, Malaysia, Maldives, Mexico, Morocco, Myanmar, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, & United Arab Emirates.
Lot, serial or model codes
Serial No. 200, 202, 204, 206, 208, 210, 212, 214, 216, 218, 220, 222, 224, 226, 228, 230, 232, 234, 238, 240, 242, 248, 250, 252, 254, 256, 258, 260, 262, 264, 266, 268, 270, 272, 274, 276, 278, 280, 282, 284, 286, 290, 292, 294, 296, 298, 300, 302, 304, 308, 312, 320, 334, 336, 338, 340, 342, 352, 354, 356, 358, 360, 362, 364, 366, 368, 370, 372, 374, 378, 380, 382, 384, 386, 388, 396, 400, 402, 406, 408, 418, 420, 422, 424, 426, 428, 432, 434, 438, 440, 442, 444, 450, 452, 454, 456, 460, 462, 464, 468, 478, 480, 482, 484, 486, 488, 490, 492, 498, 502, 504, 516, 526, 530, 536, 544, 546, 548, 550, 554, 556, 558, 560, 562, 564, 570, 572, 574, 576, 580, 582, 584, 586, 588, 590, 598, 604, 612, 614, 628, 632, 636, 640, 642, 646, 650, 654, 660, 662, 666, 668, 674, 676, 684, 686, 692, 694, 696, 698, 700, 702, 704, 706, 712, 718, 720, 726, 728, 730, 732, 734, 736, 738, 742, 744, 746, 770, 784, 794, 796, 808, 810, 814, 816, 818, 820, 824, 834, 858, 860, 864, 866, 868, 870, 874, 878, 886, 892, 314, 316, 448, 538, 594, 622, 712, 778.
510(k) numbers
not on file
PMA numbers
not on file
Product code
IEO
Firm FEI
3000210133
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register