FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BD ProbeTec" CT/GC/AC Reagent Pack, catalog #440450, labeled in part ***Becton Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152*** The BD ProbeTec" ET Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays, when tested with the BD ProbeTec ET System, use Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical swabs, male urethral swabs, and in female and male urine specimens as evidence of infection with C. trachomatis, N. gonorrhoeae, or of co-infection with both C. trachomatis and N. gonorrhoeae. Specimens may be from symptomatic or asymptomatic females and males. A separate Amplification Control is an option for inhibition testing (BD ProbeTec ET CT/GC/AC Reagent Pack). The BD ProbeTec ET CT/GC assays may be performed using either the BD ProbeTec ET System or a combination of the BD ProbeTec ET System and BD Viper" instrument.

Recall number
Z-2658-2011
Recalling firm
Becton Dickinson & Co.
Product
BD ProbeTec" CT/GC/AC Reagent Pack, catalog #440450, labeled in part ***Becton Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152*** The BD ProbeTec" ET Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays, when tested with the BD ProbeTec ET System, use Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical swabs, male urethral swabs, and in female and male urine specimens as evidence of infection with C. trachomatis, N. gonorrhoeae, or of co-infection with both C. trachomatis and N. gonorrhoeae. Specimens may be from symptomatic or asymptomatic females and males. A separate Amplification Control is an option for inhibition testing (BD ProbeTec ET CT/GC/AC Reagent Pack). The BD ProbeTec ET CT/GC assays may be performed using either the BD ProbeTec ET System or a combination of the BD ProbeTec ET System and BD Viper" instrument.
Status
Terminated
Initiated
2010-09-29
Posted
2011-06-24
Terminated
2011-06-24
Reason
In vitro diagnostic reagent packs may be underfilled and contribute to invalid quality control testing or false positive patient results.
Root cause
Process control
Action
BD Diagnostic Systems sent an "URGENT PRODUCT RECALL" letter dated September 2010 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers were instructed to discontinue use of the product and discard any remaining product. BD will replace any discarded products. Additionally, customers were requested to complete and return an attached form whether or not they have any inventory remaining, to confirm receipt of the notification. The letter instructed the customers to fax the completed form to: Attention: Regulatory Compliance at 410-316-4258. For assistance regarding replacements, contact BD Customer Service Department at 1-800-675-0908. All other inquiries should contact BD Technical Services Department at 1-800-638-8663.
Quantity
130 kits
Distribution
Worldwide Distribution--USA (nationwide) including the states of AL, CA, CO, CT, FL, GA, KY, LA, ME, MN, MO, MS, MT, NE, OH, PA, RI, TX, and WA and the countries of Australia, Canada, Europe, and Japan.
Lot, serial or model codes
Lot 0158908 exp May 31, 2011 and Lot 0158969 exp May 31, 2011
510(k) numbers
["K984631"]
PMA numbers
not on file
Product code
MKZ
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register