Philips North America Llc: EarlyVue VS30 Vital Signs Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863359

Recall number
Z-2655-2023
Recalling firm
Philips North America Llc
Product
EarlyVue VS30 Vital Signs Monitor, utilizing E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa) simultaneously, Product Code 863359
Status
Open, Classified
Initiated
2023-08-31
Posted
2023-09-29
Terminated
not on file
Reason
A software issue prevents the monitor from alarming when the high/low acoustic respiration rate (RRa) limits are violated.
Root cause
Software design
Action
An URGENT Medical Device Correction notification letter dated 8/29/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users If you use EarlyVue VS30 devices with both options E01 (Oridion Microstream Sidestream etCO2) and option R21 (Masimo Rainbow RRa): 1) Observe the EarlyVue VS30 monitor display screen for out-of-limit acoustic respiration rates (RRa) during procedures when there is simultaneous use of E01 (Oridion Microstream sidestream etCO2) and R21 (Masimo Rainbow RRa), while Respiration Source Mode is set to Auto, and the CO2 hardware module setting is switched to Standby. This notice should be passed on to all those who need to be aware within your organization or to any organization where EarlyVue VS30 with both options E01 (Oridion Microstream Sidestream etCO2) and option R21 (Masimo Rainbow RRa) devices have been transferred. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will perform software upgrade to your device. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377.
Quantity
13 total devices
Distribution
Worldwide - US Nationwide distribution in the states of MA, MI. The country of Brazil.
Lot, serial or model codes
UDI-DI: 00884838075900; Serial Numbers: CN82203600
510(k) numbers
["K190624"]
PMA numbers
not on file
Product code
DSJ
Firm FEI
3016618143
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register