FDA Medical Device Recalls (openFDA)
Philips Medical Systems (Cleveland) Inc: Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Recall number
- Z-2653-2016
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
- Status
- Terminated
- Initiated
- 2016-07-20
- Posted
- not on file
- Terminated
- 2018-07-02
- Reason
- After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
- Root cause
- Software Design Change
- Action
- Customers were sent a letter dated July 28, 2016 informing them of the problem, under what circumstances it can occur and the actions that the customer can take to minimize the effect of the problem. The letter also includes the actions planned by Philips to correct the problem. Philips plans on releasing a field change order (FCO) to correct the affected systems. If you need any further information or support concerning this issue, contact your local Philips representative or local Philips Healthcare office. For North America and Canada, contact the Customer Care Solutions Center (1 -800-722-9377).
- Quantity
- 33
- Distribution
- US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY
- Lot, serial or model codes
- 728243 Brilliance BigBore Oncology CT 7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042
- 510(k) numbers
- ["K033357"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28