FDA Medical Device Recalls (openFDA)
Medtronic Inc., Cardiac Rhythm and Heart Failure: Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1, DDBB1D4, DDBB2D1, DDBB2D4, DVBB2D4, DVBB1D1. EVERA S DDBC3D1, DDBC3D4 Product Usage: The Medtronic Evera DR dual chamber and Evera VR single chamber, implantable cardioverter defibrillator (lCD) are a family of multiprogrammable cardiac devices that monitor and regulate the patients heart rate by providing dual or single chamber rate-responsive bradycardia pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies (dual-chamber devices only).
- Recall number
- Z-2651-2016
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure
- Product
- Medtronic, Evera Implantable Cardioverter Defibrillators: Product Model EVERA XT DDBB1D1, DDBB1D4, DDBB2D1, DDBB2D4, DVBB2D4, DVBB1D1. EVERA S DDBC3D1, DDBC3D4 Product Usage: The Medtronic Evera DR dual chamber and Evera VR single chamber, implantable cardioverter defibrillator (lCD) are a family of multiprogrammable cardiac devices that monitor and regulate the patients heart rate by providing dual or single chamber rate-responsive bradycardia pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies (dual-chamber devices only).
- Status
- Terminated
- Initiated
- 2016-08-12
- Posted
- 2016-08-25
- Terminated
- 2017-01-20
- Reason
- 78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.
- Root cause
- Nonconforming Material/Component
- Action
- Consignees were sent or hand delivered on August 12, 2016 a Medtronic "Urgent Medical Device Correction" letter dated August 2016. The letter was addressed to Physician. The letter described the problem and the devices involved in the recall. The letter described a number of "Patient Management Recommendations". For questions contact your local Medtronic Representative or Medtronic Technical Services at 800-723-4636.
- Quantity
- 78 (39 US, 39 OUS)
- Distribution
- Worldwide Distribution - US Nationwide in the states of US: AZ, CA, FL, GA, IL, KY, MD, MN, NJ, NY, NC, OH, OK, PA, PR, TX. and countries of: Australia, Finland, France, Germany, Iran, Italy, Japan, Korea, Netherlands, Puerto Rico, Saudi Arabia, Spain, and United Kingdom.
- Lot, serial or model codes
- Product Name: Evera XT DR lCD. Models: DDBB1D4, DDBB1D1, DDBB2D4, DDBB2D1. Serial Numbers: BWC202738H, BWC202754H, BWB202998H, BWB203157H, BWB203167H, BWB203173H, BWB203186H, BWE601558S, BWE601571S, BWE601578S, BWE601579S, BWE601581S, BWE601589S, BWE601591S, BWE601594S, BWE601600S, BWE601605S, BWD602122S. Product Name: Evera S DR lCD. Models: DDBC3D1, DDBC3D4. Serial Numbers: BWG600597S, BWF600969S, BWF600970S, BWF600972S, BWF600973S, BWF600975S, BWF600977S, BWF600978S, BWF600979S, BWF600983S, BWF600984S, BWF600985S, BWF600987S, BWF600989S, BWF600991S, BWF600992S, BWF600996S, BWF601001S. Product Name: Evera XT VR lCD. Models: DVBB1D1, DVBB2D4. Serial Numbers: BWI201423H, BWI201436H, BW1201440H, BWI201451H, BWI201454H, BWI201462H, BWI201473H, BWJ601102S, BWJ601108S, BWJ601112S, BWJ601479S, BWJ601101S, BWJ601103S, BWJ601106S.
- 510(k) numbers
- not on file
- PMA numbers
- ["P980016S382"]
- Product code
- LWS
- Firm FEI
- 2182208
- Firm city
- Mounds View
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28