Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.

Recall number
Z-2651-2014
Recalling firm
Beckman Coulter Inc.
Product
Coulter DxH Diluent, Catalog No. 628017 Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.
Status
Terminated
Initiated
2014-08-25
Posted
2014-09-16
Terminated
2015-02-25
Reason
Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.
Root cause
Under Investigation by firm
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated August 25, 2014, to all customers that purchased the Coulter DxH Diluent. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions were instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com, or call (800) 526-7694 in the US and Canada.
Quantity
46,050 units total (40,078 units in US)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, Mexico, Panama, and Thailand. .
Lot, serial or model codes
Lot No. 3510470, 3510480, 3510490, 3510500, 3510510, 3510520, 3510530, 3510540, 3510550, 3510570, 3510580, 3510590, 3510600, 3510610, 3510620, 3510630, 3510690.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GIF
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register