Skip to the content

FDA Medical Device Recalls (openFDA)

Bovie Medical Corporation: BVX-450BR Bovie Precise 360 Handpiece 45cm, blade Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.

Recall number
Z-2648-2016
Recalling firm
Bovie Medical Corporation
Product
BVX-450BR Bovie Precise 360 Handpiece 45cm, blade Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.
Status
Terminated
Initiated
2016-07-18
Posted
not on file
Terminated
2016-12-14
Reason
The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.
Root cause
Nonconforming Material/Component
Action
Consignees were notified by written recall notice sent via certified mail or courier on July 18, 2016 to quarantine their inventory of the Precise 360 Handpiece, identify and contact their customers that have received the affected product and provide the a copy of the attached field safety notice. Customers are requested to return affected product, and return all affected product currently in their possession. All customer inventory should be reviewed, discontinue use, and return all Bovie Precise 360 handpieces to: Quality Assurance Department, Bovie Medical Corporation, 5115 Ulmerton Road, Clearwater, Florida 33760, Attn: Precise 360 Recall. Also, please write the authorization number RMA XXXX on the outside of the shipping container and use UPS account number 342562 and return via ground shipping and Fax or email the attached recall return response form to the email or fax indicated on the form as soon as possible and include a copy with the shipment. A completed recall return response form is required even if the customer has no product to return. If you have any questions about this recall or the product affected by the recall, call 727-803-8617.
Quantity
95
Distribution
Worldwide Distribution - US Nationwide in the states of: MD, GA, MO and FL and the country of Finland.
Lot, serial or model codes
Lot 0316D.
510(k) numbers
["K152570"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3007593903
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register