Skip to the content

FDA Medical Device Recalls (openFDA)

Computerized Medical Systems Inc: ERGO++ Stereotactic Radiation Treatment Planning System, ERGO++ Release 1.6.3 and 1.6.3.1 Product Usage: Used to create treatment plans for any cancer patient for who external beam radiation therapy has been prescribed. It is an accessory to linear accelerators used for radiation therapy. It is indicated for use in the planning of 3-dimensional radiation therapy.

Recall number
Z-2645-2011
Recalling firm
Computerized Medical Systems Inc
Product
ERGO++ Stereotactic Radiation Treatment Planning System, ERGO++ Release 1.6.3 and 1.6.3.1 Product Usage: Used to create treatment plans for any cancer patient for who external beam radiation therapy has been prescribed. It is an accessory to linear accelerators used for radiation therapy. It is indicated for use in the planning of 3-dimensional radiation therapy.
Status
Terminated
Initiated
2009-12-11
Posted
2011-06-23
Terminated
2014-04-01
Reason
ERGO Release 1.6.3 is overestimating the MU values.
Root cause
Software design
Action
Computerized Medical Systems, Inc. sent an "USER NOTIFICATION" dated December 11, 2009, to all affected customers. The notice identified the product, problem and suggested a workaround to minimize the risk. The notice stated that this workaround does NOT substitute the need for regular Quality Assurance. Customers are instructed to review page 39 of user manual for further explanation of warnings. Customers were asked to complete the Notice for Customer section and return to 3D Line Medical System. This issue has been resolved in Release 1.7.0 and above. No follow-up notice will be distributed to customers. For further information please contact: CMS Software (314) 812-4460 (St Louis Office)
Quantity
45
Distribution
Nationwide distribution to medical facilities in CA, MO, NC, ND, NV, NY, OH, OK, TX, and WA. Foreign distribution was made to Brazil, Cuba, Czech Republic, France, Germany, India, Italy, Japan, Malaysia, Poland, Portugal, Serbia, and Spain.
Lot, serial or model codes
ERGO++ Release 1.6.3 and 1.6.3.1
510(k) numbers
["K080601"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
1937649
Firm city
Maryland Heights
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register