Skip to the content

FDA Medical Device Recalls (openFDA)

Ventana Medical Systems, Inc.: Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD

Recall number
Z-2644-2024
Recalling firm
Ventana Medical Systems, Inc.
Product
Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Status
Open, Classified
Initiated
2024-06-27
Posted
2024-08-16
Terminated
not on file
Reason
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
Root cause
Unknown/Undetermined by firm
Action
On July 5, 2024, Ventana Medical Systems, Inc. a subsidiary of Roche Diagnostics issued a "Urgent Medical Device Correction" via UPS to affected consignees. Ventana asked consignees to take the following actions: 1. Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. 2. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. 3. Review all positive results generated with the affected lots. Follow your standard laboratory operating procedures to investigate any suspected false positive results generated with the affected lots. 4. If you performed your initial validation using an affected lot, you should revalidate. 5. Refer to the Product Replacement or Credit section on page three of this UMDC. To receive replacement product OR credit for the unexpired product currently in your inventory, use the following link to access the form: https://go.roche.com/TP-02194. 6. Complete all sections of the enclosed Recall Response form (TP-02195); fax or email it according to the instructions on the form even if you are not requesting replacement product or credit. " Share this notification with all device users within your facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred. " File this Urgent Medical Device Correction (UMDC) for future reference.
Quantity
22,839 kits
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Albania, Argentina, Armenia, Australia, Azerbaijan, Bahrain, Belarus, Brazil, Cameroon, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Georgia, Ghana, Guatemala, Honduras, Hong Kong, India, Iran, Iraq, Israel, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mali, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, New Zealand, Nicaragua, Nigeria, North Macedonia, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Sri Lanka, Taiwan, Tanzania, Thailand, T¿rkiye, Uganda, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Lot, serial or model codes
UDI: 04015630972579/J04613, J11853, J17541, J25047, J30286, K00982, K06239, K09880, K14266, K19784, K26461, and M00669
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
1000125183
Firm city
Oro Valley
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register