Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: ARTIS icono biplane. Model Number: 11327600.

Recall number
Z-2641-2026
Recalling firm
Siemens Medical Solutions USA, Inc
Product
ARTIS icono biplane. Model Number: 11327600.
Status
Open, Classified
Initiated
2026-05-20
Posted
2026-07-01
Terminated
not on file
Reason
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
Root cause
Software design
Action
Siemens notified consignees on about 05/19/2026 via letter. Consignees were instructed to ensure all users of the affected products within each organization and others who may need to be informed will receive the relevant safety information provided within the notice and comply with the recommendations therein, maintain the recall notification with product related records and until the corrective measures have been completed. Siemens will be releasing a software update to correct the issue via an Update Instruction AX018/26/S. All affected systems will be corrected by Siemens Healthineers. The service organization will contact consignees in the near future to arrange a date to perform the corrective action. Additionally, consignees were instructed to confirm and acknowledge receipt of the recall notification digitally.
Quantity
85 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model Number: 11327600. UDI-DI: 4056869063317. Serial Numbers: 184246, 184241, 184086, 184178, 184108, 184099, 184070, 184127, 184095, 184167, 184184, 184205, 184199, 184221, 184126, 184047, 184093, 184044, 184096, 184214, 184100, 184119, 184073, 184188, 184134, 184150, 184081, 184153, 184085, 184243, 184117, 184045, 184159, 184151, 184177, 184079, 184087, 184169, 184116, 184157, 184163, 184123, 184090, 184118, 184140, 184230, 184224, 184195, 184248, 184225, 184190, 184106, 184046, 184078, 184200, 184142, 184138, 184166, 184152, 184113, 184186, 184196, 184105, 184102, 184147, 184120, 184121, 184124, 184211, 184233, 184216, 184050, 184067, 184112, 184122, 184220, 184089, 184097, 184234, 184103, 184098, 184084, 184183, 184179, 184109
510(k) numbers
["K241572"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register