FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: ARTIS icono biplane. Model Number: 11327600.
- Recall number
- Z-2641-2026
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS icono biplane. Model Number: 11327600.
- Status
- Open, Classified
- Initiated
- 2026-05-20
- Posted
- 2026-07-01
- Terminated
- not on file
- Reason
- Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
- Root cause
- Software design
- Action
- Siemens notified consignees on about 05/19/2026 via letter. Consignees were instructed to ensure all users of the affected products within each organization and others who may need to be informed will receive the relevant safety information provided within the notice and comply with the recommendations therein, maintain the recall notification with product related records and until the corrective measures have been completed. Siemens will be releasing a software update to correct the issue via an Update Instruction AX018/26/S. All affected systems will be corrected by Siemens Healthineers. The service organization will contact consignees in the near future to arrange a date to perform the corrective action. Additionally, consignees were instructed to confirm and acknowledge receipt of the recall notification digitally.
- Quantity
- 85 units
- Distribution
- US Nationwide distribution.
- Lot, serial or model codes
- Model Number: 11327600. UDI-DI: 4056869063317. Serial Numbers: 184246, 184241, 184086, 184178, 184108, 184099, 184070, 184127, 184095, 184167, 184184, 184205, 184199, 184221, 184126, 184047, 184093, 184044, 184096, 184214, 184100, 184119, 184073, 184188, 184134, 184150, 184081, 184153, 184085, 184243, 184117, 184045, 184159, 184151, 184177, 184079, 184087, 184169, 184116, 184157, 184163, 184123, 184090, 184118, 184140, 184230, 184224, 184195, 184248, 184225, 184190, 184106, 184046, 184078, 184200, 184142, 184138, 184166, 184152, 184113, 184186, 184196, 184105, 184102, 184147, 184120, 184121, 184124, 184211, 184233, 184216, 184050, 184067, 184112, 184122, 184220, 184089, 184097, 184234, 184103, 184098, 184084, 184183, 184179, 184109
- 510(k) numbers
- ["K241572"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28