Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Incorporated: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Recall number
Z-2640-2026
Recalling firm
Accuray Incorporated
Product
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Status
Open, Classified
Initiated
2026-06-16
Posted
2026-07-01
Terminated
not on file
Reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Root cause
Software design
Action
Accuray issued am Urgent Field Safety Notice to it consignees via hand delivered letter on 6/16/2026. The notice explained the issue, potential risk to health and provided the following safety instructions: "This condition may occur during collimator housing drop off or pick up while the collimator housing is positioned over the Xchange table. The system will interlock, stop motion, and display a message at the operator console to contact service. If an interlock occurs: (1) remove the patient from the treatment room before troubleshooting, (2) do not initiate treatment robot motion from the in-room control pendant, and (3) contact Accuray Service to safely release the collimator housing." Accuray is correcting the instruments. For questions, please contact Accuray Customer Support by phone or email, using the Service Request form available at http://www.accuray.com/service-requests.
Quantity
48 units
Distribution
Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.
Lot, serial or model codes
All systems prior with software version 11X that utilize the Exchange Table
510(k) numbers
["K170788"]
PMA numbers
not on file
Product code
IYE
Firm FEI
3003873069
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28