FDA Medical Device Recalls (openFDA)
Howmedica Osteonics Corp.: 1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;
- Recall number
- Z-2640-2025
- Recalling firm
- Howmedica Osteonics Corp.
- Product
- 1. EXETER V40 STEM 44MM NO 2, Model/Catalog Number: 0580-1-442; 2. EXETER V40 STEM 37.5MM NO 0, Model/Catalog Number: 0580-1-352;
- Status
- Open, Classified
- Initiated
- 2025-08-28
- Posted
- 2025-09-26
- Terminated
- not on file
- Reason
- a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
- Root cause
- Under Investigation by firm
- Action
- On August 28, 2025 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions Needed: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list. 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. Under 21 CFR 803, manufacturers are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep Stryker informed of any adverse events associated with this product by emailing soprodexpreports@stryker.com. We regret any inconvenience this action may cause. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
- Quantity
- 55 units (all OUS)
- Distribution
- International distribution in the country of United Kingdom.
- Lot, serial or model codes
- 1. Model/Catalog Number: 0580-1-442, UDI-DI: 04546540153319, Lot Number: A00976; 2. Model/Catalog Number: 0580-1-352, UDI-DI: 04546540153241, Lot Number: G8754849;
- 510(k) numbers
- ["K193429"]
- PMA numbers
- not on file
- Product code
- JDI
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28