FDA Medical Device Recalls (openFDA)

Merge Healthcare, Inc.: SpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and C02, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output.

Recall number
Z-2640-2017
Recalling firm
Merge Healthcare, Inc.
Product
SpO2 connecting cable for Masimo LNCS sensor (Merge Part Number: HW-HEMO-00014), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and C02, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output.
Status
Terminated
Initiated
2016-04-04
Posted
not on file
Terminated
2019-10-15
Reason
Under certain conditions, the cables used with the Hemo system during a patient procedure could stop displaying the patient SpO2 causing a slight delay in patient care.
Root cause
Component change control
Action
Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016 via e-mail the week of 4/4/2016. If no email address was available, the letters were issued via certified mail. A second letter dated July 11, 2016 was issued to non-responders. The letter identified the affected product problem and actions to be taken.
Quantity
219 cables
Distribution
Distribution was nationwide. There was also government and military distribution. There was no foreign distribution.
Lot, serial or model codes
The following lot numbers are affected and are not specific to each cable/sensor type:   09AP8, 10AEU, 10KBN, 09CF9, 10BMN, 10MDN, 09CNS, 10FEN, 10MDN, 09KG6, 10F95, 10DUT, 09KG6, 10F95, 10DDK, 09NMN, 10BMN, 10DUT, 10A81, 10GJB, 10KBS, 09NMN, 10BMN, 10DUT, 10AEU, 10GJB, 10KBS, 10AEU, 10KBN, 10KBS, 11AGS, 11KAJ, 11CFT, 10KFP, 11N61, 11DAY, 11AGS, 11NG2, 11CFT, 11AGS, 12CNX, 11DDK, 12FMH, 12C68, 11FBS, 12GEB, 12C68, 11JBK, 12GEB, and 12FMH.
510(k) numbers
["K051212"]
PMA numbers
not on file
Product code
DQK
Firm FEI
1000116130
Firm city
Hartland
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register