FDA Medical Device Recalls (openFDA)
Beckman Coulter Inc.: Access System-GI Monitor (CA19-9), A83873D, Catalog No. 387687
- Recall number
- Z-2638-2017
- Recalling firm
- Beckman Coulter Inc.
- Product
- Access System-GI Monitor (CA19-9), A83873D, Catalog No. 387687
- Status
- Terminated
- Initiated
- 2017-05-09
- Posted
- not on file
- Terminated
- 2018-03-30
- Reason
- The Access immunoassays are susceptible to biotin interference.
- Root cause
- Device Design
- Action
- An urgent Medial Device Recall letter dated 5/9/17 was sent to customers to inform them that Beckman Coulter is initiating a field action for Access Total T3, Access Thyroglobulin, Access Free T4, and Access GI Monitor. Beckman Coulter has determined through customer feedback and an internal investigation that the four Access immunoassays listed above are susceptible to biotin interference. During interference testing the interference occurred with samples that contained 100 ng/mL of biotin. This level of biotin is greater than the maximum biotin concentration observed in the normal healthy population. Customers are informed of the impact, action, and resolution. Customers are instructed to respond within 10 days: electronically via email or manually by completing and returning the enclosed response form. Customers with any questions regarding the notice are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.
- Quantity
- 249,597 units total (9,388 units in US)
- Distribution
- Algeria, Angola, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macau, Malawi, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, Vietnam, Zambia, Zimbabwe
- Lot, serial or model codes
- all lots
- 510(k) numbers
- ["K033038"]
- PMA numbers
- not on file
- Product code
- NIG
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28