FDA Medical Device Recalls (openFDA)
North American Rescue LLC.: JUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training- Complete (Modules 1-3), REF: 85-0550; KIT, TCCC MODULE ONE - SKILLS, REF: 85-0519; KIT, MULTI MISSION EXPEDITIONARY RESPONSE - MMERK, REF: 85-2420; KIT, TCCC SKILLS/IFAK COMBO, REF: 85-0740, KIT, TCCC/TECC MODULE ONE SKILLS REFILL, REF: 85-1900
- Recall number
- Z-2637-2025
- Recalling firm
- North American Rescue LLC.
- Product
- JUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training- Complete (Modules 1-3), REF: 85-0550; KIT, TCCC MODULE ONE - SKILLS, REF: 85-0519; KIT, MULTI MISSION EXPEDITIONARY RESPONSE - MMERK, REF: 85-2420; KIT, TCCC SKILLS/IFAK COMBO, REF: 85-0740, KIT, TCCC/TECC MODULE ONE SKILLS REFILL, REF: 85-1900
- Status
- Open, Classified
- Initiated
- 2025-07-01
- Posted
- 2025-09-26
- Terminated
- not on file
- Reason
- Device used for unilateral or bilateral occlusion of blood flow to the lower limbs has a potential manufacturing defect that may compromise the structural integrity of the device and device performance during use.
- Root cause
- Nonconforming Material/Component
- Action
- On 7/1/2025, recall notices were emailed to customers who were asked to do the following: 1) Quarantine the junctional emergency treatment tools. 2) Discontinue use and distribution. 3) Complete and return the response form via email to recalls@narescue.com In addition, if you have further distributed affected products, identify your customers and notify them of this recall. Your notification to your customers may be enhanced by including a copy of the recall notice. Instruct your customers to return affected product to you and then you will coordinate return with the recalling firm. This recall should be carried out to the user level. If you have questions contact the firm's customer service at 1-888-689-6277
- Quantity
- 552
- Distribution
- Worldwide - US Nationwide distribution in the states of SC, OK, ID, KY, VA, NV, IL, TX, CA, FL, LA, NY, LA, ME, PA, AR, AZ, AL, DE, MA, MS, HI, WA, MI, GA, NJ, UT, MD, OR, NC, NC, AL and the countries of KOR, JPN, BWA, SAU, BEL.
- Lot, serial or model codes
- JETT: REF: 30-0088 UDIs: (01)00842209100002(11)240426(10)0212241, (01)00842209100002(11)240515(10)817221, (01)00842209100002(11)240530(10)817222, (01)00842209100002(11)240618(10)817222, (01)00842209100002(11)240701(10)817222, (01)00842209100002(11)240712(10)1213231. REF: 30-0088: Lots: 021224-1, 81722-1, 81722-2, 121323-1. Kits: REF/UDI-DI/Lot: 85-0550/00842209110711/85-0550091824, 85-0550090424, 85-0550030725, 85-0550052925, 85-0550052325, 85-0550082324, 85-0550080624, 85-0550062624, 85-0550052124, 85-0550111424, 85-0550102924; 85-0519/00842209110742/85-0519090324, 85-0519081924; 85-2420/00842209113910/85-2420041725, 85-2420042125, 85-2420100324; 85-0740/00842209110599/85-0740042925, 85-0740050625, 85-0740052925, 85-0740081224, 85-0740082924, 85-0740011325, 85-0740122324, 85-0740121724, 85-0740112024, 85-0740110424, 85-0740102524; 85-1900/00842209110810/85-1900082124, 85-1900031125
- 510(k) numbers
- ["K123194"]
- PMA numbers
- not on file
- Product code
- DXC
- Firm FEI
- 3005034064
- Firm city
- Greer
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as