Elekta, Inc.: MOSAIQ(R) Oncology Information System, Versions 2.50.05 and higher Product Usage: MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Users can configure MOSAIQ for Medical Oncology use, Radiation Oncology use, or the two together. It lets users: Assemble electronic patient charts and treatment plans, order diagnostic tests, and prescribe medications. Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology. Import, view, annotate, adjust, enhance, manage and archive images. Compare radiation treatment plans and evaluate dose coverage. Design leaf plans for operation with radiotherapy treatment machines that have multileaf collimators. Make sure radiation treatment plans imported from treatment planning systems agree with treatment machine constraints. MOSAIQ reads actual settings from the treatment machine through the machine communication interface. It compares these settings with predefined values. If a mismatch occurs between the planned values and the actual machine settings, the system warns the user. View reference images to setup treatment. MOSAIQ refers to predefined settings to help treatment machine setup, and communicates patient and machine setup instructions. Record actual delivered radiation values in an electronic chart to track treatment. MOSAIQ is not intended for use in diagnosis. Medical oncology dose calculation functions are designed for use with patients 18 years or older only.

Recall number
Z-2633-2017
Recalling firm
Elekta, Inc.
Product
MOSAIQ(R) Oncology Information System, Versions 2.50.05 and higher Product Usage: MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Users can configure MOSAIQ for Medical Oncology use, Radiation Oncology use, or the two together. It lets users: Assemble electronic patient charts and treatment plans, order diagnostic tests, and prescribe medications. Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology. Import, view, annotate, adjust, enhance, manage and archive images. Compare radiation treatment plans and evaluate dose coverage. Design leaf plans for operation with radiotherapy treatment machines that have multileaf collimators. Make sure radiation treatment plans imported from treatment planning systems agree with treatment machine constraints. MOSAIQ reads actual settings from the treatment machine through the machine communication interface. It compares these settings with predefined values. If a mismatch occurs between the planned values and the actual machine settings, the system warns the user. View reference images to setup treatment. MOSAIQ refers to predefined settings to help treatment machine setup, and communicates patient and machine setup instructions. Record actual delivered radiation values in an electronic chart to track treatment. MOSAIQ is not intended for use in diagnosis. Medical oncology dose calculation functions are designed for use with patients 18 years or older only.
Status
Terminated
Initiated
2017-06-21
Posted
not on file
Terminated
2018-07-11
Reason
Retention of Outdated Information in an Order Set. In the Care Plan Library, double clicking to add or delete a Component/Procedure to an Order Set does not trigger the update of the associated Care Plans.
Root cause
Software design
Action
Elekta sent an Urgent Important Field Safety Notification letter dated June 2017. The notice informs users of the specific product and version numbers affected by the issue, and any work arounds that can be used to avoid the issue. Customers are instructed that a copy of the Important Field Safety Notice is to be kept with the most current labelling and all personnel working with the product should be made aware of the content of the letter. The customers are instructed to complete and return the Important Field Safety Notice Acknowledgement form to their local Elekta representative as soon as possible but no longer than 30 days from receipt.
Quantity
355 units
Distribution
Worldwide Distribution - US Nationwide and the countries of ANTIGUA AND BARBUDA, AUSTRALIA, AUSTRIA, BAHRAIN, CANADA, CHEZ REPUBLIC, GERMANY, HUNGARY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWEDEN, AND UK
Lot, serial or model codes
None
510(k) numbers
["K123230"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register