Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO PACK, Part Number DYNDA1431A; 2) CVC INSERT BUNDLE 16CM, Part Number ECVC3260; 3) CENTRAL LINE INSERTION KIT, Part Number ECVC7440; 4) CENTRAL LINE INSERTION KIT, Part Number ECVC7445; 5) CENTRAL LINE INSERTION KIT W/20CM CATH, Part Number ECVC7445A; 6) VANTEX 7FR 3L 20CM CVC BUNDLE US, Part Number ECVC8045; 7) CVC DOUBLE LUMEN 16CM, Part Number ECVC8105

Recall number
Z-2632-2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO PACK, Part Number DYNDA1431A; 2) CVC INSERT BUNDLE 16CM, Part Number ECVC3260; 3) CENTRAL LINE INSERTION KIT, Part Number ECVC7440; 4) CENTRAL LINE INSERTION KIT, Part Number ECVC7445; 5) CENTRAL LINE INSERTION KIT W/20CM CATH, Part Number ECVC7445A; 6) VANTEX 7FR 3L 20CM CVC BUNDLE US, Part Number ECVC8045; 7) CVC DOUBLE LUMEN 16CM, Part Number ECVC8105
Status
Open, Classified
Initiated
2024-04-08
Posted
2024-08-27
Terminated
not on file
Reason
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Root cause
Nonconforming Material/Component
Action
Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
Quantity
1,679,067 units in total
Distribution
Worldwide distribution.
Lot, serial or model codes
DYNDA1431A, Lot Number 2019021190; ECVC3260, Lot Number 2019091850; ECVC3260, Lot Number 2019120950; ECVC7440, Lot Number 2021071950; ECVC7440, Lot Number 2021092150; ECVC7445, Lot Number 2021052850; ECVC7445, Lot Number 2021062250; ECVC7445A, Lot Number 2022012850; ECVC7445A, Lot Number 2023010390; ECVC8045, Lot Number 2023031590; ECVC8105, Lot Number 2023112190
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRP
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register