Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Inc.: Servo-s Ventilator Systems; Article number 66 40 440. Intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory insufficiency; these are only to be used by healthcare providers in hospitals or healthcare facilities and for in-hospital transport. These are not to be used with any anesthetic agents.

Recall number
Z-2629-2011
Recalling firm
Maquet Inc.
Product
Servo-s Ventilator Systems; Article number 66 40 440. Intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory insufficiency; these are only to be used by healthcare providers in hospitals or healthcare facilities and for in-hospital transport. These are not to be used with any anesthetic agents.
Status
Terminated
Initiated
2011-04-20
Posted
2011-06-23
Terminated
2013-09-17
Reason
Through customer complaints, Maquet has been informed that SERVO-i ventilators have stopped to ventilate and have generated three different technical error codes.
Root cause
Other
Action
Maquet Inc. sent an URGENT DEVICE SAFETY ALERT letter dated April 20, 2011, to all affected customers via Federal Express. The letter identified the product, the problem and the action needed to be taken by the customer. The letter informs the user of the issue and provides additional instruction and how the existing manual should be updated. Customers are asked to fill out the attached response form and fax it to 1-973-807-9210. The following warning will be included in all user's manuals delivered with new ventilators: 'Warning ! The Servo-i Ventilator System is not intended to be used during radiotherapy, since this may cause system malfunction. Warning! The Servo-s Ventilator System is not intended to be used during radiotherapy, since this may cause system malfunction.' Additionally the warning stated above will be included in the format of a sticker with the yearly preventive maintenance kit for Servo-i and SErvo-s to update exiting User's manuals for the installed base of Servo-i and SErvo-s ventilators. A maintenance report form will also be included in the kit which serves as an acknowledgement from the user that the sticker has been added to an affected ventilator unit's operating manual. For further questions, please call (973) 709-7660.
Quantity
17039 units worldwide; 842 units US
Distribution
Worldwide Distribution - USA (nationwide)
Lot, serial or model codes
Article number 66 40 440; all units
510(k) numbers
not on file
PMA numbers
not on file
Product code
CBK
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register