Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Co.: BACTEC" FX Instrument-Top, catalog #441385. The BD BACTEC" FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.

Recall number
Z-2627-2011
Recalling firm
Becton Dickinson & Co.
Product
BACTEC" FX Instrument-Top, catalog #441385. The BD BACTEC" FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.
Status
Terminated
Initiated
2010-08-10
Posted
2011-06-23
Terminated
2011-07-05
Reason
In vitro diagnostic instrumentation may exhibit random operational errors that could delay effective patient treatment.
Root cause
Software design
Action
BD Diagnostics Systems sent an "URGENT PRODUCT RECALL" letter dated August 2010 to all affected customers. The letter describes the product, problem and action to be taken. The letter described the type of error messages the customers may experience. All affected units will be scheduled for on site software update as well as a disk drive replacement. A BD field service engineer will contact the customers to schedule the on site visit. Customers experiencing issues prior to the scheduled visit or have questions should contact the BD Diagnostics Service Department at 800-638-8663.
Quantity
907units
Distribution
Worldwide Distribution--USA (nationwide) including countries of Australia, Belgium, Canada, China, Colombia, New Zealand, Hong Kong, Japan, Korea, Philippines, Singapore, Taiwan, and Thailand.
Lot, serial or model codes
Catalog # 441385-Serial numbers: FT-0001 to FT-0906 and FT-0922
510(k) numbers
not on file
PMA numbers
not on file
Product code
MDB
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register