Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Spine: Stryker Aria Implant Inserter, REF 48758500, This instrument is part of the Aria product brand. It is used for Implant insertion into the disc space. The Aria Inserter PEEK Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights, and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation, ergonomically shaped anterior edges, and flat posterior edges. This device does not sustain or support life.

Recall number
Z-2626-2014
Recalling firm
Stryker Spine
Product
Stryker Aria Implant Inserter, REF 48758500, This instrument is part of the Aria product brand. It is used for Implant insertion into the disc space. The Aria Inserter PEEK Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights, and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation, ergonomically shaped anterior edges, and flat posterior edges. This device does not sustain or support life.
Status
Terminated
Initiated
2014-07-07
Posted
2014-09-11
Terminated
2017-05-08
Reason
Stryker has initiated the recall of its Aria Standard and Specialty Implant Inserters due to complaints of product fracturing.
Root cause
Device Design
Action
Stryker Spine sent an "Urgent Medical Device Removal Notification/Customer Response" form dated July 7, 2014 to all their affected customers via Fedex. Customers were instructed to examine their inventory and hospital locations to identify the product. Fax a copy of the Customer Response Form to 855-632-9049 or email it to Spine-RegulatoryAction@Stryker.com. For questions regarding this recall call 201-760-8298.
Quantity
64 units
Distribution
Worldwide Distribution - USA including OR, AZ, TN, MA, MD, GA, FL, NY, IA, IN, TX, MI, PA, UT, CA, MN, MO, OH, KS, OK, MT, FL, AL, CO, TN, and WI and Internationally to Australia.
Lot, serial or model codes
Catalog #48758500, Lot #'s 107400, 10G906 and 110214;
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register