FDA Medical Device Recalls (openFDA)
Stryker Howmedica Osteonics Corp.: GMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used with the components of the Howmedica Osteonics Modular Replacement System , the Howmedica Osteonics Modular Rotating Hinge Knee System and/or the Howmedica Osteonics Kinematic Rotating Hinge Knee System in situations where replacement of extensive bone loss in the femur and/or proximal tibia is required. The Proximal Femoral Module and Distal Femoral Module of the GMR may be used together, or with the components of the MRS, in replacement of the total femur.
- Recall number
- Z-2624-2010
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Product
- GMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used with the components of the Howmedica Osteonics Modular Replacement System , the Howmedica Osteonics Modular Rotating Hinge Knee System and/or the Howmedica Osteonics Kinematic Rotating Hinge Knee System in situations where replacement of extensive bone loss in the femur and/or proximal tibia is required. The Proximal Femoral Module and Distal Femoral Module of the GMR may be used together, or with the components of the MRS, in replacement of the total femur.
- Status
- Terminated
- Initiated
- 2005-05-05
- Posted
- 2010-09-29
- Terminated
- 2010-10-01
- Reason
- Mislabeled package: Package labeled as GMRS Extension Piece 60 mm, lot ZYCOC1 actually contains a GMRS Extension Piece 70 mm extension piece.
- Root cause
- Labeling mix-ups
- Action
- Stryker issued Important Market WIthdrawal letters dated May 5, 2005 to customers and distributors. The letters identified the affected product, the labeling issue, and actions to be taken by the customer or distributor. Customers were instructed to identify and quarantine affected product, which will be reconciled and returned by the Stryker sales representative. Distributors were instructed to identify affected product, reconcile all product on an attached Product Accountability Form to be faxed to Stryker at 201 831-4825 within two days of receipt of the letter. The affected product was to be returned using the enclosed pre-paid mailing label. Contact the Stryker sales representative to re-order product that is being returned. Stryker can be contacted at 201 831-5825.
- Quantity
- 5 units
- Distribution
- Nationwide Distribution: USA, including the states of AZ, FL, MD, and PA.G
- Lot, serial or model codes
- Catalog number: 6495-6-060; Lot code LZYOC1
- 510(k) numbers
- ["K023087"]
- PMA numbers
- not on file
- Product code
- KRO
- Firm FEI
- 3003070421
- Firm city
- Mahwah
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28