FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics, INc.: VITROS XT Chemistry Products TBIL-ALKP Slides
- Recall number
- Z-2623-2023
- Recalling firm
- Ortho-Clinical Diagnostics, INc.
- Product
- VITROS XT Chemistry Products TBIL-ALKP Slides
- Status
- Open, Classified
- Initiated
- 2023-08-05
- Posted
- 2023-09-21
- Terminated
- not on file
- Reason
- Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
- Root cause
- Under Investigation by firm
- Action
- An "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letter dated 08/17/2023 was issued to impacted consignees. Consignees are provided instructions to utilize when reviewing sample results and any result between 600-1500 U/L should be diluted using a 10x dilution and retested. A confirmation of receipt of the recall letter should be completed and returned by August 25, 2023. Please forward this notification if the affected product was distributed outside your facility. Save this notification with your documentation or post this notification by each VITROS 250/350/FS 5,1/4600/5600/XT 3400/XT 7600 System until the issue has been resolved. If your laboratory has experienced this issue with the product and has not already done so, please report the occurrence to your local Ortho Care Technical Solutions Center. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
- Quantity
- 52,456 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
- Lot, serial or model codes
- Product Code: 6844296 UDI: (10758750031801) GEN #16; Expiry Date Range 01 Jul 2023 - 01 Sep 2023 GEN #17; Expiry Date Range 01 Nov 2023 - 01 Feb 2024 GEN #18; Expiry Date Range 01 Apr 2024 - 01 Aug 2024 GEN #19; Expiry Date Range 01 Aug 2024 - 01 Jan 2025
- 510(k) numbers
- ["K190807"]
- PMA numbers
- not on file
- Product code
- CIG
- Firm FEI
- 1000305840
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28