Covidien LLC: Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid or for blunt dissection of soft tissue(s) and structures, or as an aid in controlling minor intraoperative bleeding
- Recall number
- Z-2622-2017
- Recalling firm
- Covidien LLC
- Product
- Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid or for blunt dissection of soft tissue(s) and structures, or as an aid in controlling minor intraoperative bleeding
- Status
- Terminated
- Initiated
- 2017-05-09
- Posted
- not on file
- Terminated
- 2019-02-13
- Reason
- Cotton tip of the device may disengage due to insufficient adhesive
- Root cause
- Process control
- Action
- Medtronic consignees were notified via Federal Express or certified mail on May 9, 2017, and the letter informs customers of the potential for the tip of the device to disengage from the device and the actions they should take. Accounts requested to discontinue use, quarantine and return product. Customers are requested to complete and acknowledge receipt of the Urgent Field Corrective Action Notice. Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2. For questions regarding this recall call 203-492-5000.
- Quantity
- 194,851
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Lot Numbers: M5K04X to M7A01X Exp Date: 2020-October through 2022-January
- 510(k) numbers
- ["K931962"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 1219930
- Firm city
- North Haven
- Firm state
- CT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28