Covidien LLC: Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid or for blunt dissection of soft tissue(s) and structures, or as an aid in controlling minor intraoperative bleeding

Recall number
Z-2622-2017
Recalling firm
Covidien LLC
Product
Covidien Endo Peanut 5mm Auto Suture Blunt Dissector Item Code: 173019 The instrument has indications in endoscopic surgery for swabbing small amounts of fluid or for blunt dissection of soft tissue(s) and structures, or as an aid in controlling minor intraoperative bleeding
Status
Terminated
Initiated
2017-05-09
Posted
not on file
Terminated
2019-02-13
Reason
Cotton tip of the device may disengage due to insufficient adhesive
Root cause
Process control
Action
Medtronic consignees were notified via Federal Express or certified mail on May 9, 2017, and the letter informs customers of the potential for the tip of the device to disengage from the device and the actions they should take. Accounts requested to discontinue use, quarantine and return product. Customers are requested to complete and acknowledge receipt of the Urgent Field Corrective Action Notice. Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2. For questions regarding this recall call 203-492-5000.
Quantity
194,851
Distribution
Nationwide Distribution
Lot, serial or model codes
Lot Numbers: M5K04X to M7A01X Exp Date: 2020-October through 2022-January
510(k) numbers
["K931962"]
PMA numbers
not on file
Product code
GCJ
Firm FEI
1219930
Firm city
North Haven
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register