Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew Inc: smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Recall number
Z-2621-2023
Recalling firm
Smith & Nephew Inc
Product
smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
Status
Open, Classified
Initiated
2023-08-31
Posted
2023-09-20
Terminated
not on file
Reason
The firm conducting a recall due to a complaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 4N LT implant instead of a LEGION NARROW PSvOXIN SZ 5N LT as described on the product label.
Root cause
Labeling Change Control
Action
Smith & Nephew issued an Urgent Medical Device3 Recall Notice to its consignees on 08/31/2023 via letter and email. The notice explained the problem and potential risk and requested the consignee quarantine the affected product. Distributors were directed to notify their customers. Firm is seeking return of the affected product.
Quantity
8 units
Distribution
OH, CT, TX, NC, CA, NE, NV
Lot, serial or model codes
UDI/DI: 00885556234006, Lot Number: 23EM09878
510(k) numbers
["K112941","K230653"]
PMA numbers
not on file
Product code
JWH
Firm FEI
3005551626
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register