FDA Medical Device Recalls (openFDA)

Windstone Medical Packaging, Inc.: A M S, ALIGNED MEDICAL SOLUTIONS, Thyroid FNA Pack-RX, Pack Number AMS17729, surgical convenience kit

Recall number
Z-2620-2026
Recalling firm
Windstone Medical Packaging, Inc.
Product
A M S, ALIGNED MEDICAL SOLUTIONS, Thyroid FNA Pack-RX, Pack Number AMS17729, surgical convenience kit
Status
Open, Classified
Initiated
2026-05-21
Posted
2026-07-01
Terminated
not on file
Reason
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
Root cause
Material/Component Contamination
Action
AMS ALIGNED MEDICAL SOLUTIONS issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 5/21/2026 via email. The notice explained the issue, potential risk, and requested the following: "AMS requests that you undertake the following activities: 1. Immediately check your inventory for the recalled products listed above and place them under quarantine. 2. Due to the manufacturing process any finished sterilized packs remaining in inventory, or at your facility will require over labeling for all products affected by this recall. Packs will be labeled with a sticker attached to each pack. This sticker will read: RECALL NOTICE CARDINAL HEALTH announced the recall of 5110 Webcol" Large Alcohol Prep Pad contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall. " At the time the kit is opened for use any Cardinal Alcohol Prep Pad should be identified and set aside. " The recalled Cardinal Alcohol Prep Pad should be rendered unusable to protect against inadvertent use and disposed of pursuant to the medical waste policies in effect at your institution. 3. Complete the attached Recall Reply Form listing all inventory of affected product by lot number and quantity. 4. If you have further distributed this product, please identify your distribution partners, and notify them at once of the product recall. Your notification to your customers should include a copy of this notification letter. 5. Please complete the Recall Reply Form within 5 business days and return via email to fieldcorrectiveaction@alignedmedical.com even if you have no affected product on hand." For questions or concerns: fieldcorrectiveaction@alignedmedicalsolutions.com.
Quantity
200 units
Distribution
US Nationwide distribution in the states of PA, GA, CA, WA.
Lot, serial or model codes
UDI/DI none, Lot Codes: 237663, 237901, 237957, 238041, 238084, 238085, 238146, 238197, 238589, 239157, 239216, 239286, 239337, 239386, 239943, 240269, 240633, 240705, 240903, 240904, 241034, 241223, 241224, 241313.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OGR
Firm FEI
1000125955
Firm city
Billings
Firm state
MT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register