FDA Medical Device Recalls (openFDA)
Stryker Communications: Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
- Recall number
- Z-2620-2025
- Recalling firm
- Stryker Communications
- Product
- Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034
- Status
- Open, Classified
- Initiated
- 2025-07-15
- Posted
- 2025-09-24
- Terminated
- not on file
- Reason
- Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.
- Root cause
- Component change control
- Action
- On 7/15/2025, recall notices were mailed or hand delivered to customers who were asked to do the following: Inform individuals within your organization who need to be aware of this action. 2. A Stryker representative will be in contact with your account to schedule a field service visit at your earliest convenience to complete the following: a. Deinstall impacted Oculan light(s) and reinstall with SLX lights to ensure your operating room stays in use. b. Upon product remediation and availability, Stryker will return to reinstall Oculan within your operating room(s). Your local Stryker representative will provide timing updates as they are available. 3. Complete and return the Business Reply Form to CommunicationsRA1@stryker.com 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
- Quantity
- 40
- Distribution
- US Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY, IL, MN.
- Lot, serial or model codes
- UDI-DI: 07613327296167 and 07613327651317 REF/Serial Numbers: CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155. P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSY
- Firm FEI
- 3002975342
- Firm city
- Flower Mound
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28