Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Neuromodulation Corporation: WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320

Recall number
Z-2617-2024
Recalling firm
Boston Scientific Neuromodulation Corporation
Product
WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Status
Open, Classified
Initiated
2024-07-17
Posted
2024-08-14
Terminated
not on file
Reason
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
Root cause
Software design
Action
On 7/17/24, "URGENT MEDICAL DEVICE ADVISORY" Notices were sent to Physicians/Surgeons, Hospitals, Healthcare Professionals informing them of the following: 1) Review any patient reports of undesired sensations indicative of potential device reset behavior during IPG charging and report these observations to Boston Scientific. 2) An in-field IPG firmware update is available for the patient s device. This will eliminate the possibility of a coincident routine device system check during IPG charging, thus preventing a potential system reset. Append the patients medical record accordingly if a firmware update is completed. 3) Patient letter is enclosed with this communication. This letter is provided for sharing with your patient and/or to be included within the patient s medical record. The patient letter can also be accessed on the Boston Scientific Product Advisories page via this link: www.bostonscientific.com/en-US/pprc/product-advisories.html 4) Post this information in a visible location near the product(s) to ensure information is easily accessible to all users. 5) Distributor customers should notify all customers that have been shipped/sold affected product about the correction. 6) Complete and return the acknowledgement form to BSCFieldActionCenter@bsci.com Any adverse events or quality concerns associated with use of this product should be reported to the firm via email at BSN.ComplaintCallCenter@bsci.com The patient letter advises the following: discuss this letter and the device software update with your doctor and continue to follow their instructions. If you have any additional questions after contacting your doctor, please contact the firm's Support at 1-833-327-8324 (Option 2) or email to: bsn.tech.support@bsci.com.
Quantity
77,674
Distribution
Worldwide Distribution: US (nationwide) including states of: TN, PA, WA, SC, MN, OH, NC, AK, NY, IL, CA, NJ, KY, IA, MO, UT, FL, IN, LA, AL, MS, MD, WI, GA, VA, TX, AZ, KS, OR, CO, CT, ID, AR, OK, MA, MI, ND, NE, SD, WV, NV, HI, NH, DC, MT, ME, VT, WY, DE, NM. OUS (foreign) countries of: Canada, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary, Ireland, Israel, Italy, Lebanon, Luxembourg, Netherlands, Norway, Poland, Qatar, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkey, Ukraine, United Arab Emirates, Portugal, Japan, Australia, New Zealand, Singapore, Argentina, Brazil, Colombia, Puerto Rico
Lot, serial or model codes
REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365SC12160/8714729985082/100188 - 758920, M365SC12320/8714729985099/100635 - 759404
510(k) numbers
not on file
PMA numbers
["P030017"]
Product code
LGW
Firm FEI
3006630150
Firm city
Valencia
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28