Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics Inc. (Navilyst Medical Inc.): Xcela PICC with PASV Valve Technology/Kit 5 F-55cm; (Maximum Flow Rate 3.5 mL/Sec)-indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, and nutrients; the sampling of blood; and for power injection of contrast media. Catalog Number: 60M701995

Recall number
Z-2617-2021
Recalling firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Product
Xcela PICC with PASV Valve Technology/Kit 5 F-55cm; (Maximum Flow Rate 3.5 mL/Sec)-indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, and nutrients; the sampling of blood; and for power injection of contrast media. Catalog Number: 60M701995
Status
Terminated
Initiated
2017-10-17
Posted
not on file
Terminated
2022-06-14
Reason
Mislabeled: The product is labeled as 5F (55cm; Maximum Flow Rate 3.5 mL/Sec) and the product is actually 4F (55cm; Maximum Flow Rate 3.5 mL/Sec).
Root cause
Error in labeling
Action
Angiodynamics issued Recall Notifications Packages via Federal Express to 1 International Medical Facility on 2017-10-17. Letter states reason for recall, health risk and action to take: IMMEDIATELY " Stop using the product subject to recall. " Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). " Segregate this product in a secure location for return to AngioDynamics, Inc. " Forward a copy of this recall notification to all sites to which you have distributed affected product. Complete and return the Reply Verification Tracking Form. " If affected product is located in your institution, please can AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product. " Promptly complete, sign and return the enclosed Reply Verification Tracking Form (even if you do not have any product to return); following the directions on this page and the Reply Verification Tracking Form. o Email Reply Verification Tracking Form (preferred): recall@angiodynamics.com o Fax Reply Verification Tracking Form: Ann: PICC Kit Product Description Recall Coordinator Fax number 1 800-782-1357.
Quantity
4 boxes (5/each)= 20 eaches
Distribution
Canada
Lot, serial or model codes
Lot Number: 5210162  (UDI): H96560M7019951
510(k) numbers
not on file
PMA numbers
not on file
Product code
N/A
Firm FEI
3017892510
Firm city
Glens Falls
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register