Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Mitek, Inc., a Johnson & Johnson Co.: DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft

Recall number
Z-2617-2017
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Product
DePuy Mitek LATARJET EXPERIENCE Coracoid Top Hat Drill Product Code: 288202 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Status
Terminated
Initiated
2017-05-12
Posted
not on file
Terminated
2018-10-03
Reason
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Root cause
Component design/selection
Action
DePuy Mitek issued recall letter dated May 12, 2017 advising of the problem and requesting return of the product. A Response Form is to be completed and returned. Questions: contact your DePuy Synthes Mitek Sales Consultant or Carolyn Somerville, DePuy Mitek, Inc. (via telephone: 508.828.3647 Fax: 508.828.3762 or via email: DPYUS-MitekFieldActions@its.jnj.com).
Quantity
184 units
Distribution
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
Lot, serial or model codes
GTN: 01)10886705026807 Lot codes: 16C01, 16E03, 16R01, 17A01, 17B01
510(k) numbers
["K083096","K091694","K110763"]
PMA numbers
not on file
Product code
HWC
Firm FEI
1221934
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register