Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH

Recall number
Z-2613-2025
Recalling firm
Medline Industries, LP
Product
Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH
Status
Open, Classified
Initiated
2025-07-31
Posted
2025-09-26
Terminated
not on file
Reason
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Root cause
Under Investigation by firm
Action
On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.
Quantity
7913 eaches
Distribution
Domestic US distribution nationwide. International distribution pending. No international distribution.
Lot, serial or model codes
Item Number/UDI-DI D087031RH 10197344019557 LOTS EP240731 EP240814 EP240826 EP240903 EP240916 EP241003 EP241014 EP241028 EP241104 EP241112 EP241202 EP241209A EP241219 EP241224 EP250114 EP250116 EP250204 EP250219
510(k) numbers
["K232130"]
PMA numbers
not on file
Product code
OWQ
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register