Terumo Cardiovascular Systems Corporation: CDI TM 101 Monitor, CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System, The CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System is intended for use during cardiopulmonary bypass procedures where you need continuous monitoring of the blood's oxygen saturation, hematocrit and hemoglobin levels.
- Recall number
- Z-2610-2011
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- CDI TM 101 Monitor, CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System, The CDI TM 101 Hematocrit/Oxygen Saturation Monitoring System is intended for use during cardiopulmonary bypass procedures where you need continuous monitoring of the blood's oxygen saturation, hematocrit and hemoglobin levels.
- Status
- Terminated
- Initiated
- 2011-04-28
- Posted
- 2011-06-22
- Terminated
- 2012-11-06
- Reason
- Terumo discovered through its own testing that the presence of electromagnetic interference (EMI) in the operating room may affect the accuracy of the values displayed by the CDI"101 Hematocrit/Oxygen Saturation Monitoring System: hematocrit (HCT), hemoglobin (Hgb) and Oxygen Saturation.
- Root cause
- Device Design
- Action
- Terumo Cardiovascular Systems, Corp issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 28, 2011, to all affected customers. The letter identifed the product, the problem and the action needed to be taken by the customer. After a solution has been verified, Terumo CVS will implement a correction for all affected units. Customers were instructed to review this Medical Device Correction notice, Assure that all users are aware of this notice. Confirm receipt of this and to fax back the Customer Response form. For further question please contact Terumo CVS Customer Service at 1-800-521-2818, Monday - Friday, 8 AM - 6 PM EST.
- Quantity
- 622
- Distribution
- Worldwide Distribution -- USA ( nationwide) and the countries of Colombia, Vietnam, India, Indonesia, Pakistan, Canada, Chile, Vietnam, Chile, Hong Kong, United Arab Emirates, Singapore, Australia, Malaysia, Singapore, Taiwan, Belgium, Thailand, Vietnam and India
- Lot, serial or model codes
- Lot Number 1070-1839
- 510(k) numbers
- ["K902654"]
- PMA numbers
- not on file
- Product code
- DRY
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28