Skip to the content

FDA Medical Device Recalls (openFDA)

Ab Sciex: AB SCIEX API 3200MD" LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543.

Recall number
Z-2609-2014
Recalling firm
Ab Sciex
Product
AB SCIEX API 3200MD" LC/MS/MS System with software: MultiQuantMD 3.0In vitro diagnostic to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens Part Number 5024543.
Status
Terminated
Initiated
2014-07-15
Posted
2014-09-08
Terminated
2014-10-01
Reason
Software- AB SCIEX API 3200MD LC/MS/MS SYSTEM with MultiQuantMD 3.0 software may provide incorrect data for patient diagnosis
Root cause
Software design
Action
AB SIEX issued letter on 7/15/14, notifying customers to implement this field correction, a temporary fix to be implemented immediately. Additionally, a software update is in development and is expected to be available in approximately 8 weeks. Upon availability, AB SCIEX will send a DVD with instructions on how to install the new software update. At that time, outlined temporary actions will no longer be necessary. If you have any questions regarding this recall, potential hazards or actions to be taken please feel free to contact AB SCI EX at +1 289 982 2531. Please confirm receipt of this letter by signing and faxing back the attached Acknowledgement Form.
Quantity
6 units
Distribution
FL, NC, OK, PA, TX
Lot, serial or model codes
Serial Numbers: BV20031308, BV20081311, BV20051310, BV20071311, BV20101401, BV20131401
510(k) numbers
not on file
PMA numbers
not on file
Product code
DOP
Firm FEI
3008935545
Firm city
Framingham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28