FDA Medical Device Recalls (openFDA)
Philips North America Llc: Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154
- Recall number
- Z-2608-2023
- Recalling firm
- Philips North America Llc
- Product
- Patient Information Center iX with Software Version Number 4.x in the following: (1) Patient Information Center iX Product Number: 866389; (2) Patient Information Center iX Expand Product Number: 866390; (3) PIC iX Essentials Product Number: 867093; (4) PIC iX Essentials Expand Product Number: 867154
- Status
- Open, Classified
- Initiated
- 2023-08-17
- Posted
- 2023-09-19
- Terminated
- not on file
- Reason
- Patient Information Center (PIC iX) Release 4.x Surveillance Crash Caused by Intel Graphics Driver Error-with a blank screen and subsequently require a manual reboot of the hardware to restart and continue central patient monitoring. Potential for a delay in the detection of a change or deterioration in the condition of one or more patients
- Root cause
- Under Investigation by firm
- Action
- Philips issued Urgent field Safety Notice to to U.S. customers via certified mail on 8/17/23. Philips Market Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If central monitoring is disrupted and blank screen displays: 1) Initiate local downtime procedures to evaluate patients at bedside and confirm local monitoring devices continue to monitor. 2) Manually reboot PIC iX to continue central monitoring This notice should be passed on to all those who need to be aware within your organization or to any organization where PIC iX 4.x devices have been transferred. 5. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule a visit from a Philips Field Service Engineer who will install a software patch to your device. The patch will add defensive mitigations to force an automatic resumption of central patient monitoring without manual intervention. If you need any further information, please contact your local Philips representative, or call 1-800-722-9377.
- Quantity
- 960 units
- Distribution
- Nationwide Foreign: Austria Bahrain Belgium Canada Czech Republic Denmark Finland France Germany Greece India Ireland Italy Japan Jordan Kenya Kuwait Lebanon Lithuania Netherlands Norway Poland Portugal Qatar Romania Slovakia South Africa South Korea Spain Sweden Sweden Switzerland United Kingdom Utd.Arab.Emir. Uzbekistan
- Lot, serial or model codes
- UDI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 Software Version Number 4.x
- 510(k) numbers
- ["K153702"]
- PMA numbers
- not on file
- Product code
- MHX
- Firm FEI
- 3016618143
- Firm city
- Cambridge
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28